Performances visuelles, glistening et opacification capsulaire postérieure : 3 ans après l’implantation d’un nouvel implant hydrophobe - 06/07/18

Posterior capsule opacification, glistenings and visual outcomes: 3 years after implantation of a new hydrophobic IOL

Doi : 10.1016/j.jfo.2017.11.022 
C. Chassain a, , C. Chamard b
a Clinique Beausoleil, 119, avenue de Lodève, 34070 Montpellier, France 
b CHU de Montpellier, 191, avenue du Doyen-Gaston-Giraud, 34295 Montpellier cedex 5, France 

Auteur correspondant.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Résumé

But

Évaluation des performances visuelles, de l’incidence du glistening et de l’opacification capsulaire postérieure après l’implantation d’un nouvel implant intraoculaire hydrophobe jaune dans une série de cas consécutifs.

Design

Étude rétrospective.

Méthodes

Cette étude comprend une série d’yeux opérés de la chirurgie de la cataracte et implantés avec l’implant monofocal PODEye (PhysIOL SA, Liège, Belgique) entre mai et novembre 2011. Ils sont comparés pour la détermination de la stabilité du matériau (glistening) avec des patients implantés avec des implants intraoculaires Acrysof (Alcon, Fort Worth, États-Unis) qui ont eu une consultation de routine entre janvier et août 2015. La sécurité de l’implant est évaluée à l’aide des résultats de l’acuité visuelle, du centrage et de la médication postopératoire. L’incidence et la sévérité des glistenings ont été évaluées subjectivement avec la lampe à fente (absent=0 ; modéré=1 ; dense2). Le suivi est de 3 ans.

Résultats

L’étude comprend 66 yeux. L’acuité visuelle moyenne après l’implantation du PODEye était de de 0,98±0,06 à une semaine(n = 66); de 1,00±0,01 à 3 semaines (n= 64); de 0,99±0,02 à 3 mois (n = 59) et de 0,98±0,05 (n = 50) en valeur décimale à 1 an, l’implant a montré un comportement stable dans toutes les directions. Aucun glistening de grade 2 n’a été observé dans le groupe G-Free/PODEye alors qu’il était significativement présent dans le groupe Acrysof. Un seul œil sur 43 yeux suivis à 3 ans de recul a eu besoin d’une capsulotomie YAG à 37 mois. Soixante-six pour cent des implants présentaient une opacification capsulaire postérieure de grade 0 (absent=0 ; modéré=1 ; dense2) à 3 ans.

Conclusions

La lentille intraoculaire PODEye est sûre. Aucun glistening de grade 2 n’a été observé dans cet implant à filtre de lumière bleue. L’occurrence d’opacification capsulaire postérieure était faible à trois ans.

El texto completo de este artículo está disponible en PDF.

Summary

Purpose

To evaluate the visual outcomes, the incidence of glistenings and posterior capsule opacification in a series of consecutive eyes with a new blue light-filtering hydrophobic acrylic intraocular lens (IOL)

Setting

Service d’ophtalmologie, clinique Beausoleil, avenue de Lodève, Montpellier.

Design

Retrospective, cross-sectional study.

Methods

This study included a series of 66 eyes that underwent routine cataract surgery with PODEye (PhysIOL SA, Liège, Belgium) monofocal IOL implantation between May 2011 and November 2011. These eyes were compared for glistenings with patients implanted with Acrysof material IOL (Alcon, Fort Worth, USA) who had a routine postoperative examination between January 2015 and August 2015. The safety of the IOL was assessed by visual acuity outcomes, as well as rotational stability, centration and postoperative medication. The incidence and severity of glistenings were evaluated with the slit lamp. Glistenings and PCO were graded subjectively (0=absent; 1=moderate; 2=dense). Follow-up was 3 years.

Results

The mean visual acuity after PODEye implantation was 0.98±0.05 (n=50) in decimal notation at 1 year, 0.98±0.06 at 1 week, (n=66), 1.00±0.01 at 3 weeks (n=64) and 0.99±0.02 at 3 months (n=59). The IOL displayed stable positioning in all axes. Glistenings of grade 2 were absent in the G-Free/PODEye group, whereas they were significantly present in the Acrysof group. One out of 43 eyes at 3-year follow-up underwent YAG capulotomy at 37 months. Sixty-six percent of eyes had no PCO (grade 0) at 3 years.

Conclusions

The PODEye IOL was safe and stable. Glistenings grade 2 or more were not found with this blue light-filtering hydrophobic acrylic IOL. PCO occurrence was low at three years.

El texto completo de este artículo está disponible en PDF.

Mots clés : Implant hydrophobe, Glistening, Opacification capsulaire postérieure, IOL sécurité, Chirurgie de la cataracte

Keywords : Hydrophobic material, Glistening, Posterior capsule opacification, IOL stability, Cataract surgery


Esquema


 Le texte a fait l’objet d’une première présentation lors de l’ASCRS 2012 sous le titre « 1-year follow-up after implantation of 100 IOLs with a new hydrophobic material. ». Il s’agissait d’un rapport intérimaire.


© 2018  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 41 - N° 6

P. 513-520 - juin 2018 Regresar al número
Artículo precedente Artículo precedente
  • Perception des lésions oculaires cécitantes de la drépanocytose par les praticiens hospitalo-universitaire de Dakar
  • A.M. Wane, A.M. Ka, J.P. Diagne, G. Mouen Bengue, A.S. Sow, P.A. Ndoye Roth, E.A. Ba, M.E. De Medeiros, H. Kane, J.M. Ndiaye, H.M. Diallo, P.A. Ndiaye, M.R. Ndiaye
| Artículo siguiente Artículo siguiente
  • Exophtalmie bilatérale pseudo-tumorale révélant un syndrome hyper-IgG4 orbitaire
  • J. Brour, H. Sassi, W. Koubaa Mahjoub, M. Ouederni, A. Hassairi, A. Chedly Debbiche, M. Chéour

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.