Reduction of patient radiation dose with a new organ based dose modulation technique for thoraco-abdominopelvic computed tomography (CT) (Liver dose right index) - 11/07/18

Doi : 10.1016/j.diii.2018.01.013 
M. Fillon a, S. Si-Mohamed a, b, , P. Coulon c, A. Vuillod d, P. Klahr e, L. Boussel a, b
a Radiology Department, Centre Hospitalier Universitaire, 69000 Lyon, France 
b CREATIS, UMR CNRS 5220, Inserm U1044, University Lyon 1 Claude-Bernard, 69000 Lyon, France 
c CT Clinical Science, Philips, 69000 Suresnes, France 
d Centre Hospitalier Universitaire, Radiophysic Department, 69000 Lyon, France 
e CT Clinical Science, Philips, Cleveland, USA 

Corresponding author. CHU cardiologique Louis-Pradel, département d’imagerie cardiovasculaire, 59, boulevard Pinel, 69500 Bron, France.CHU cardiologique Louis-Pradel, département d’imagerie cardiovasculaire, 59, boulevard Pinel, 69500 Bron, France.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Purpose

To evaluate the radiation dose reduction, image quality and diagnostic confidence with thoraco-abdominopelvic computed tomography (TACT) using a new organ based dose modulation system (liver dose right index [Liver DRI]), compared to TACT using a standard automatic exposure control adjusting mA according to attenuation.

Methods

A total of 37 patients who had two TACT examinations on 2 different CT scanners, one using standard automatic exposure control (combination of a DoseRight automatic current selection and Z modulation) and one using Liver DRI were included. There were 19 men and 18 women with a mean age of 67.6±11.7 (SD) years (range: 36–85 years) For each patient, volume CT dose index (CTDIvol), size-specific dose estimates (SSDE) and signal-to-noise ratio (SNR) were evaluated at each anatomic level (lung, breast, liver and pelvis area) for each protocol. Two radiologists assessed independently image quality, artifacts and diagnostic confidence.

Results

The radiation dose decreased significantly using Liver DRI compared to standard automatic exposure control on the total scan length, lung, breast and pelvis area, with a significant CTDIvol reduction of 27% (P=0.0001), 23% (P=0.0002), 24% (P=0.0002) and 31% (P=0.0001), respectively; and a significant SSDE reduction of 23% (P=0.0001), 28% (P=0.0001), 23% (P=0.0002) and 29% (P=0.0001), respectively. No significant SNR reductions were observed in all measured tissues at the level of the aortic arch, celiac trunk and iliac bifurcation, except in the muscle (P=0.0013) and fat tissue (P=0.0052) at the level of the ureteral meatus. No significant differences were noted between both protocols in overall image quality, artifacts and diagnostic confidence with an excellent inter observer agreement between radiologists (Kappa values of 0.83, 0.85 and 0.88, respectively).

Conclusion

Liver DRI organ based dose modulation technique allows significant dose reduction compared to standard automatic exposure control while preserving diagnostic image quality in all thoraco-abdominopelvic areas.

El texto completo de este artículo está disponible en PDF.

Keywords : Patient dose optimization, Automatic exposure control, Radiation exposure, Comparative study, Signal-to-noise ratio

Abbreviations : ACS, AEC, CNR, CT, CTDI, DRI, ROI, SNR, SSDE, TACT, WED


Esquema


© 2018  Soci showét showé françaises de radiologie. Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 99 - N° 7-8

P. 483-492 - juillet 2018 Regresar al número
Artículo precedente Artículo precedente
  • Perfusion MR imaging at 3-Tesla: Can it predict tumor grade and histologic necrosis rate of musculoskeletal sarcoma?
  • P.A. Gondim Teixeira, A. Renaud, S. Aubert, M. Ben Haj Amor, Y.-M. Robin, A. Cotten, L. Ceugnart
| Artículo siguiente Artículo siguiente
  • Suspicious liver nodule in chronic liver disease: Usefulness of a second biopsy
  • V. Cartier, A. Crouan, M. Esvan, F. Oberti, S. Michalak, B. Gallix, O. Seror, A. Paisant, V. Vilgrain, C. Aubé

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.