Reverse shoulder arthroplasty for proximal humerus fractures: Is the glenoid implant problematic? - 27/09/18

Doi : 10.1016/j.otsr.2018.06.008 
Arthur Lignel a, , Julien Berhouet b, Marc-Alexandre Loirat a, Philippe Collin c, Hervé Thomazeau a, David Gallinet d, Pascal Boileau e, Luc Favard d
a CHU Pontchaillou, 35000 Rennes, France 
b CHU de Tours, 37000 Tours, France 
c CHP de Saint-Grégoire, 35760 Saint-Grégoire, France 
d Upper Limb Orthopedic Surgery University, 25000 Besançon, France 
e CHU de Nice, 06000 Nice, France 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Introduction

Reverse shoulder arthroplasty (RSA) is a key tool in the orthopedic trauma surgeon's arsenal, especially when faced with a proximal humerus fracture in older patients. However, few studies have focused on the glenoid side of RSA in this indication as the implant is placed in a generally healthy scapula.

Hypothesis

Glenoid implants for RSA after trauma are well positioned and do not often cause complications.

Material and methods

Retrospective multicenter study of 513 patients who underwent RSA because of a proximal humerus fracture. The mean follow-up was 55 months. Radiographs were used to assess the height and tilt of the glenoid implant, along with the development of scapular notching or loosening. The clinical outcomes were determined based on the Constant score.

Results

At the last follow-up, 44% of shoulders had scapular notching, 7% of which were severe (stages 3–4). This notching was progressive, with two resulting in loosening. The rate of severe notching was higher in patients with a high glenoid implant (62.5% vs. 42.3%, p=0.03) or glenosphere with superior tilt (58.3% vs. 37.8%, p=0.02). Nine patients had confirmed loosening and 63 had potential loosening. This was more common in cases with superior tilt (9.3% vs. 0.4%, p<0.001). Patients with a high glenoid implant had a lower Constant score (57 vs. 45, p<0.001). There fewer cases of severe notching when a lateralized glenoid implant was used (0% vs. 7%, p<0.05) and/or the humeral implant had a smaller neck-shaft angle (implants<155°: 3% vs. implants at 155°: 8.5%, p=0.03).

Discussion and conclusion

Glenoid loosening and severe scapular notching are related to poor positioning and/or incorrect orientation of the glenosphere. Implant selection is important, as there is little to no notching when less-angled humeral implants and lateralized glenoid implants are used.

Level of evidence

IV.

El texto completo de este artículo está disponible en PDF.

Keywords : Shoulder joint, Orthopedic Surgical Procedure, Traumatology, Joint Prosthesis


Esquema


© 2018  Publicado por Elsevier Masson SAS.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 104 - N° 6

P. 773-777 - octobre 2018 Regresar al número
Artículo precedente Artículo precedente
  • Accuracy of patient-specific guided implantation of the glenoid component in reversed shoulder arthroplasty
  • O. Verborgt, A.I. Hachem, K. Eid, K. Vuylsteke, M. Ferrand, P. Hardy
| Artículo siguiente Artículo siguiente
  • Reverse shoulder arthroplasty in recent proximal humerus fractures
  • Jean-David Werthel, François Sirveaux, Damien Block

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.