MRI detection of radiographically occult fractures of the hip and pelvis in the elderly: Comparison of T2-weighted Dixon sequence with T1-weighted and STIR sequences - 01/03/19

Doi : 10.1016/j.diii.2018.11.008 
B. Heynen, C. Tamigneaux, V. Pasoglou, J. Malghem, B. Vande Berg, T. Kirchgesner
 Department of Medical Imaging, Cliniques Universitaires Saint-Luc, Université Catholique de Louvain, Institut de Recherche Expérimentale et Clinique (IREC), 1200 Brussels, Belgium 

Corresponding author. Cliniques universitaires Saint-Luc, Department of Medical Imaging, 10, avenue Hippocrate, 1200 Brussels, Belgium.Cliniques universitaires Saint-Luc, Department of Medical Imaging10, avenue HippocrateBrussels1200Belgium

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Purpose

To compare the diagnostic performance of T2-weighted Dixon, T1-weighted and Short-Tau Inversion Recovery (STIR) MR images for the detection of radiographically occult fractures (ROF) of the hip and pelvis in elderly patients after low-energy trauma.

Materials and methods

A total of 22 patients older than 50 years with suspected ROF after low-energy trauma was prospectively included. There were 9 men and 13 women, with a mean age of 80.9 years±12.5 (SD) (range: 52–100 years). T2-weighted Dixon, T1-weighted and STIR MR images were analyzed by 3 independent radiologists blinded to the clinical data and the results of other imaging examinations. Readers separately assessed each series of images for the presence of fractures on a per bone analysis. Diagnostic performance of each reader was compared for Dixon and non-Dixon sequences using contingency tables and McNemar test. Interobserver agreement was evaluated according to the Fleiss-Cuzick's kappa statistics.

Results

The sensitivity of the Dixon sequence in the detection of ROF ranged from 90.9% (20/22; 95% CI: 70.8–98.9%) to 100% (22/22; 95% CI: 84.6–100%). The sensitivities of the non-Dixon sequences in the detection of ROF ranged from 95.5% (21/22; 95% CI: 77.2–99.9%) to 100% (22/22; 95% CI: 84.6–100%). For each reader, there were no statistical differences between combined Dixon and combined non-Dixon images for the detection of ROF (P=0.12, 0.99 and 0.99). Interobserver agreement with T2-weighted Dixon water-only images was significantly lower than that with the STIR sequence (0.70–0.79 vs. 0.87–0.93).

Conclusion

T2-weighted Dixon may be a second-rate alternative to T1-weighted and STIR sequences for the detection of ROF of the hip and pelvis in elderly patients.

El texto completo de este artículo está disponible en PDF.

Keywords : Bone fracture, Hip, Pelvis, MRI, Dixon sequence


Esquema


© 2018  Société française de radiologie. Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 100 - N° 3

P. 169-175 - mars 2019 Regresar al número
Artículo precedente Artículo precedente
  • Virtual reconstruction of paranasal sinuses from CT data: A feasibility study for forensic application
  • P. Gach, L. Tuchtan-Torrents, C. Delteil, P. Adalian, M.D. Piercecchi, L.C. Ebert, G. Gorincour
| Artículo siguiente Artículo siguiente
  • Convolutional neural network evaluation of over-scanning in lung computed tomography
  • M. Colevray, VM. Tatard-Leitman, S. Gouttard, P. Douek, L. Boussel

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.