Suscribirse

Influence of iterative reconstruction and dose levels on metallic artifact reduction: A phantom study within four CT systems - 26/04/19

Doi : 10.1016/j.diii.2018.12.007 
J. Greffier , A. Larbi, J. Frandon, P.A. Daviau, J.P. Beregi, F. Pereira
 Department of Radiology, CHU Nîmes, Université Montpellier, Medical Imaging Group Nîmes, EA 2415, 30029 Nîmes cedex, France 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Purpose

To compare metallic artifact reduction (MAR) algorithms proposed by four vendors according to the delivered dose and iterative level using a phantom study.

Methods

Four CT systems (Revolution GSI®, Ingenuity Elite®, Somatom Edge®, and Aquilion Prime®) equipped with MAR algorithms (Smart MAR®, O-MAR®, iMAR®, and SEMAR®) were compared. Acquisitions were performed with CIRS model 062M Phantom containing a titanium rod core insert using 120kV and two dose levels (3 and 7mGy). Images were reconstructed with and without MAR algorithms using standard “soft tissue” kernel for filtered back projection (FBP) and intermediary iterative level. Artifact propagation was assessed by counting the number of pixels containing a HU outside a defined threshold interval (>100HU and<−80 HU). Artifact correction was evaluated by computing the differences between images with and without MAR.

Results

Accuracy of NCT values increased significantly using MAR algorithms, IR, and high dose levels (P<0.001). Image noise reduced −31±15 (SD) % (range: −50%;−14%) with Smart-MAR®, −28±3 (SD) % (range −31%; −25%) with O-MAR®, −32±7 (SD) % (range: −40%; −24%) with iMARCN®, −52±8 (SD) % (range: −60%; −42%) with iMARTH® and −29±6 (SD) % (range: −37%; −23%) with SEMAR®. The number of pixels outside the threshold interval was also reduced using MAR algorithms. Each MAR algorithm corrected in distinct patterns, with satisfactory artifact correction for all MAR algorithms.

Conclusion

This study demonstrates that artifact correction using MAR algorithms differs according to the main manufacturers, although corrections are satisfactory for all systems. Corrections also improved by using IR and increasing the dose level.

El texto completo de este artículo está disponible en PDF.

Keywords : Metallic artifact reduction (MAR), Computed tomography (CT), Image quality, Iterative reconstruction

Abbreviations : CT, FBP, iMAR®, IR, MAR, O-MAR®, ROI, SEMAR®, Smart-MAR®


Esquema


© 2019  Société française de radiologie. Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 100 - N° 5

P. 269-277 - mai 2019 Regresar al número
Artículo precedente Artículo precedente
  • Sensorineural hearing loss in patients with vestibular schwannoma correlates with the presence of utricular hydrops as diagnosed on heavily T2-weighted MRI
  • M. Eliezer, G. Poillon, C. Maquet, A. Gillibert, J. Horion, J.-P. Marie, J.-P. Guichard, N. Magne, A. Attyé
| Artículo siguiente Artículo siguiente
  • Temporal evaluation of the microwave ablation zone and comparison of CT and gross sizes during the first month post-ablation in swine lung
  • H. Kodama, E. Ueshima, K. Howk, S.W. Lee, J.P. Erinjeri, S.B. Solomon, G. Srimathveeravalli

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.