Noise assessment across two generations of iterative reconstruction algorithms of three manufacturers using bone reconstruction kernel - 26/11/19

Doi : 10.1016/j.diii.2019.08.003 
J. Greffier , J. Frandon, A. Larbi, D. Om, J.P. Beregi, F. Pereira
 EA 2415, Department of Radiology, Medical Imaging Group Nîmes, University of Montpellier, CHU Nîmes, boulevard du Professeur Robert Debré, 30029 Nîmes cedex, France 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Purpose

To compare the noise-magnitude and noise-texture obtained using strong kernel across two generations of iterative reconstruction (IR) algorithms proposed by three manufacturers.

Materials and methods

Five computed tomography (CT) systems equipped with two generations of IR algorithm (hybrid/statistical IR [H/SIR] or full/partial model-based IR [MBIR]) were compared. Acquisitions on Catphan 600 phantom were performed at 120kV and three dose levels (CTDIvol: 3, 7 and 12mGy). Raw data were reconstructed using standard “bone” kernel for filtered back projection and one iterative level of two generations of IR algorithms. Contrast-to-noise ratio (CNR) was computed using three regions of interest placed semi-automatically: two placed in the low-density polyethylene and Teflon inserts and another placed on the solid water. Noise power spectrum (NPS) was computed to assess the NPS-peak and noise-texture.

Results

CNR was significantly greater in MBIR compared to H/SIR algorithms for all CT systems (P<0.01). CNR were improved on average from H/SIR to MBIR of 36±14% [SD] (range: 24–57%) for GE-Healthcare, 109±19 [SD] % (range: 89–139%) for Philips Healthcare and 42±5 [SD] % (range: 36–47%) for Siemens Healthineers. The mean NPS peak decreased from H/SIR to MBIR by −41±6 [SD] % (range: −47–−35%) for GE Healthcare, −79±3 [SD] % (range: −82–−76%) for Philips Healthcare and −52±2 [SD] % (range: −54–−51%) for Siemens Healthineers systems. NPS spatial frequencies were greater with MBIR than with H/SIR for Philips Healthcare (20 ± 2 [SD] %; range: 19–22%) and for Siemens Healthineers (9±5 [SD] %; range: 4–14%) but lower for GE Healthcare (−17±3 [SD] %; range: −14–−20%).

Conclusion

Using bone kernel with recent MBIR algorithms reduces the noise-magnitude for all CT systems assessed. Noise texture shifted towards high frequency for Siemens Healthineers and Philips Healthcare but the opposite for GE Healthcare.

El texto completo de este artículo está disponible en PDF.

Keywords : Multidetector computed tomography, Image quality enhancement, Iterative reconstruction, Optimization, Noise power spectrum


Esquema


© 2019  Société française de radiologie. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 100 - N° 12

P. 763-770 - décembre 2019 Regresar al número
Artículo precedente Artículo precedente
  • Sedation and analgesia in interventional radiology: Where do we stand, where are we heading and why does it matter?
  • F.H. Cornelis, E. Monard, M.-A. Moulin, E. Vignaud, F. Laveissiere, M. Ben Ammar, M. Nouri-Neuville, M. Barral, B. Lombart
| Artículo siguiente Artículo siguiente
  • Impact of an ultrasound-guided radiofrequency ablation training program on the outcomes in patients with hepatocellular carcinoma
  • R. Takai Takamatsu, A. Okano, G. Yamakawa, K. Mizukoshi, H. Obayashi, M. Ohana

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.