Suscribirse

Éléments de compréhension de l’ostéolyse de la coracoïde après l’intervention de Latarjet : étude en éléments finis comparant trois méthodes de fixation - 30/01/20

Understanding the causes behind coracoid graft osteolysis in Latarjet procedure: Finite element analysis and comparison of three fixation methods

Doi : 10.1016/j.rcot.2019.11.023 
Nazmi Bülent Alp a, , Oğuz Doğan b, Tufan Gürkan Yılmaz b, Onur Can Kalay b, Abdellah Ait Moussa c, Fatih Karpat b, Morshed Khandaker c, Gokhan Akdag d
a Department of Orthopedics and Traumatology, Uludag University School of Medicine, Bursa, Turquie 
b Department of Mechanical Engineering, Uludag University, Bursa, Turquie 
c Department of Engineering and Physics, University of Central Oklahoma, Edmond, États-Unis 
d Department of Orthopedics and Traumatology, İstanbul Beylikduzu State Hospital, Istanbul, Turquie 

Auteur correspondant. Bursa Uludag University, Medical Faculty, Orthopedics and Traumatology Department, 16059 Gorukle, Bursa, Turquie.Bursa Uludag University, Medical Faculty, Orthopedics and Traumatology DepartmentGorukle, Bursa16059Turquie

Abstract

Background

Latarjet is one of the most common surgical procedure performed on patients with shoulder instability with osseous defects, some complications include coracoid bone graft osteolysis, osteoarthritis, graft detachment, and malpositioning were previously reported. Several studies investigated potential causes of graft osteolysis but still, it remains a crucial area of investigation. We aim to use finite element analysis to examine the potential correlations between three modes of fixation methods used in the Latarjet procedure (screw, wedge plate, and endobutton), and the coracoid graft osteolysis.

Hypothesis

Finite element analysis tested the hypothesis that there is a linear relationship between the compression stress on graft which was generated by fixation methods used in the Latarjet and the coracoid graft osteolysis.

Material and methods

Boundary conditions and inhomogeneous material properties were carefully assigned within the material of the scapula and coracoid interface. For the screw and wedge plate fixations, an applied torque in the range of 1–1.5Nm was used to characterize the surgeon's insertion torque during the surgical operation, while a 100N compressive force was selected for the endobutton fixation.

Results

Relatively lesser stress magnitudes were observed with endobutton fixation method rather than screw and the wedge plate fixation. Statistical analyses revealed significant differences between the groups (p<0.05).

Discussion

Excessive compressive stresses within the coracoid graft regions may be responsible for osteolysis due to negative effects over biological factors such as blood flow. Our study emphasizes the importance of taking into account the fixation method while performing the Latarjet procedure. We concluded that the mode of fixation used within the Latarjet procedure has a correlation on the coracoid graft osteolysis.

Level of evidence

I.

El texto completo de este artículo está disponible en PDF.

Keywords : Latarjet, Osteolysis, Finite element analysis, Wedge plate, Endobutton



 Cet article peut être consulté in extenso dans la version anglaise de la revue Orthopaedics & Traumatology: Surgery & Research sur Science Direct (sciencedirect.com) en utilisant le DOI ci-dessus.


© 2019  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 106 - N° 1

P. 29 - février 2020 Regresar al número
Artículo precedente Artículo precedente
  • Nouveaux facteurs prédictifs de la rupture du subscapulaire : angle de coraco-petite tubérosité humérale, angle et hauteur de la tubérosité inférieure
  • Joong-Bae Seo, Sung-Joon Kim, Hee Jung Ham, Kwon Young Kwak, Jaesung Yoo
| Artículo siguiente Artículo siguiente
  • Évaluation objective de la qualité du sommeil chez les patients souffrant d’une rupture de la coiffe des rotateurs
  • Chase B. Ansok, Lafi S. Khalill, Stephanie Muh

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

¿Ya suscrito a @@106933@@ revista ?

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.