Est-il possible de placer des vis de verrouillage dans les parties ulnaires des plaques verrouillées palmaire à distance du tendon du long fléchisseur du pouce ? Étude cadavérique - 06/02/20

Ulnar facet locking screw locations of volar locking plates placed without flexor pollicis longus tendon contact: A cadaver study

Doi : 10.1016/j.rcot.2019.12.012 
Yuki Kikuchi a, Kotaro Sato a, , Yoshikuni Mimata a, Kenya Murakami a, Gaku Takahashi a, b, Minoru Doita a
a Department of orthopaedic surgery, Iwate medical university, 1-1-1, Idaidori, Yahaba-cho, Shiwa-gun, Iwate 028-3694, Japan 
b Department of critical care medicine, Iwate medical university, 1-1-1, Idaidori, Yahaba-cho, Shiwa-gun, Iwate 028-3694, Japan 

Auteur correspondant.
En prensa. Pruebas corregidas por el autor. Disponible en línea desde el Thursday 06 February 2020
This article has been published in an issue click here to access

Abstract

Background

The volar locking plate (VLP) system provides stable fixation and is widely used for distal radius fractures. Studies have shown that irritation of the implant prominence is a risk factor for flexor tendon rupture, especially of the flexor pollicis longus (FPL). Contact between VLPs and flexor tendons should be avoided. We aimed to investigate the ulnar facet locking screw locations of various VLPs placed without FPL tendon contact in cadaver wrists.

Hypothesis

We hypothesized that ulnar facet locking screws would be in the vicinity of the subchondral bone when the plate is placed in the most distal position without FPL tendon contact.

Materials and methods

The study assessed two variable-angles and four fixed locking plates. We placed each plate in six different cadavers, resulting in 36 different plate-cadaver combinations. Plates were placed in the most distal position without FPL tendon contact. We drilled the most ulnar hole (hole A) and the second ulnar hole (hole B) of the distal row. All drilling procedures were performed using a specific jig for each fixed locking plate. For variable-angle locking plates, we drilled with a fixed jig for each plate. We obtained lateral radiographs when the drill penetrated the dorsal cortex and measured the distance between the drill and the articular surface.

Results

With regard to hole A, the mean distances between the drill and the center of the articular surface were 2.6–5.2mm for the four fixed locking plates and 4.9–5.6mm for the two variable-angle locking plates. With regard to hole B, the mean distances between the drill and the center of the articular surface were 3.8–5.9mm for the four fixed locking plates and 5.5–5.9mm for the two variable-angle locking plates.

Discussion

When clinicians place a VLP without FPL tendon contact, the distance between the ulnar facet screws and the center of the articular surface is over 3mm in most cases. Surgeons should select variable-angle drilling for strong articular support when using variable-angle locking plates.

Level of evidence

III, diagnostic Level.

El texto completo de este artículo está disponible en PDF.

Keywords : Distal radius fracture, Volar locking plate, Flexor pollicis longus, Tendon rupture, Anatomic study



 Cet article peut être consulté in extenso dans la version anglaise de la revue Orthopaedics & Traumatology: Surgery & Research sur Science Direct (sciencedirect.com) en utilisant le DOI ci-dessus.


© 2019  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

¿Ya suscrito a @@106933@@ revista ?

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.