Effect of tin filter-based spectral shaping CT on image quality and radiation dose for routine use on ultralow-dose CT protocols: A phantom study - 29/05/20

Doi : 10.1016/j.diii.2020.01.002 
J. Greffier , F. Pereira, A. Hamard, T. Addala, J.P. Beregi, J. Frandon
 Department of Radiology, CHU Nîmes, Univ Montpellier, Medical Imaging Group Nîmes, EA 2415, 30029 Nîmes, France 

Corresponding author. Department of Radiolgy, CHU de Nîmes, Medical Imaging Group Nîmes, EA 2415, Bd Prof Robert Debré, 30029 Nîmes Cedex 9, France.Department of Radiolgy, CHU de Nîmes, Medical Imaging Group Nîmes, EA 2415Bd Prof Robert DebréNîmes Cedex 930029France

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Purpose

The purpose of this study was to assess the impact of tin filter (TF) on X-ray beam quality, image quality and radiation dose and its suitability for routine use for chest and lumbar-spine/pelvis-hip ultralow-dose (ULD) CT examination protocols.

Materials and methods

The X-ray beam quality was determined by measuring the half-value layer (HVL) and calculating the mean weighted energy for 100, 120, 150kVp (using standard filtration) and for 100 and 150kVp using TF (Sn100kVp and Sn150kVp, respectively). Acquisitions were performed on a phantom at four dose levels for each previously definedkVp. The mean attenuation (NCT), noise-power-spectrum (NPS) and task-based transfer function (TTF) were computed. The detectability index (d’) was computed to model the detection of two lesions in spine and pelvic/hip examination and two for chest exploration. Image quality and detectability using a TF were assessed for two routinely used ULD protocols.

Results

The HVL and mean weighted energy increased using a TF for the same tube voltage. Using a TF for the same tube voltage changed NCT for bone and acrylic inserts, decreased the NPS peak without changing the NPS spatial frequency and increased the TTF values. The d’ values were improved using a TF and with the dose increase. d’ values of all modeled lesions were improved using Sn100kVp and Sn150kVp for the lumbar-spine/pelvis-hip and chest ULD protocols except for sclerotic bone lesion using Sn150kVp.

Conclusion

The use of TF increases the X-ray beam quality and improves the image quality characteristics in phantom images, thus appearing as a promising tool for reducing dose and/or improving the image quality of ULD protocols.

El texto completo de este artículo está disponible en PDF.

Keywords : Multidetector computed tomography, Image enhancement, Radiation dose, Ultra low dose protocol.

Abbreviations : ADMIRE, CT, CTDIvol, d’, ESF, HVL, IR, LSF, NPS, NPWE, ROI, TF, TTF, ULD


Esquema


© 2020  Société française de radiologie. Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 101 - N° 6

P. 373-381 - juin 2020 Regresar al número
Artículo precedente Artículo precedente
  • Sequential treatment by polidocanol and radiofrequency ablation of large benign partially cystic thyroid nodules with solid components: Efficacy and safety
  • Y. Lin, P. Li, Y.-P. Shi, X.-Y. Tang, M. Ding, Y. He, B. Zhai
| Artículo siguiente Artículo siguiente
  • Quantitative signal intensity ratios to distinguish between subfascial lipoma and atypical lipomatous tumor/well-differentiated liposarcoma using short-tau inversion recovery (STIR) MRI
  • R. Donners, A.H. Krieg, D. Baumhoer, D.T. Boll, D. Harder

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.