Ablation des clous d’allongement intramédullaires PRECICE Stryde cassés - 30/06/21
Removal of broken PRECICE Stryde intramedullary lengthening nails
This article has been published in an issue click here to access
Abstract |
This paper describes two different techniques for removal of broken Precice Stryde intramedullary bone lengthening nails, which, unlike trauma nails, are solid containing mechanical components. Consequently, surgeons face unique challenges when these implants brake within medullary canal. Here, we present our surgical approach for removal of three broken implants. In one patient (46kg), both Ø10mm femoral Stryde implants (max. weight allowance 68kg) broke through the proximal locking screw hole preoperatively on the right side and intraoperatively on the left side (413 and 504 days after index surgery, respectively). The third Ø11.5 femoral nail broke through the area containing the magnet (55kg patient, 325 days after index surgery). Level of evidence: IV.
El texto completo de este artículo está disponible en PDF.Keywords : Bone lengthening, Intraoperative complications, Postoperative complications, Stryde, Hardware removal
Abbreviations : IMLN
☆ | Cet article peut être consulté in extenso dans la version anglaise de la revue Orthopaedics & Traumatology: Surgery & Research sur Science Direct (sciencedirect.com) en utilisant le DOI ci-dessus. |