Force and aspiration on catheters utilized in the ADAPT technique in acute ischemic stroke: A bench top analysis - 12/05/22

Doi : 10.1016/j.neurad.2019.03.013 
Mickey Smith a, b, , Jonathan Pace a, b, Connie Ju b, Yin C. Hu a, b
a University Hospitals Cleveland Medical Center, The Neurological Institute, Department of Neurosurgery, 11100 Euclid Avenue, Cleveland, 44106 OH, USA 
b Case Western Reserve University School of Medicine, 2109 Adelbert Road, Cleveland, 44106 OH, USA 

Corresponding author at: University Hospitals Cleveland Medical Center, The Neurological Institute, Department of Neurosurgery, 11100 Euclid Avenue, Cleveland, 44106 OH, USA.University Hospitals Cleveland Medical CenterThe Neurological InstituteDepartment of Neurosurgery11100 Euclid AvenueClevelandOH44106USA

Abstract

Introduction

Given the high morbidity and mortality of stroke, there remains a demand for techniques that provide rapid and safe intervention while improving time to recanalization. The direct aspiration first pass technique (ADAPT) uses force and aspiration for clot removal without the aid of separators or retrievers. In this study, we compare the force and aspiration qualities of commercially available catheters.

Methods

Four different catheters with varying inner diameters were set up in a bench top model to test catheter tip pressure and flow rate. Catheter tip pressure was measured by attaching the catheter to a vacuum pressure gauge and an aspiration pump. The flow rate was calculated by measuring the volume of room temperature water aspirated through each catheter over a given time.

Results

The Microvention Sofia catheter generated the greatest tip force (21.32 g), and the Stryker AXS Catalyst 6 catheter generated the smallest tip force (15.88 g). The Penumbra ACE 068 catheter and Medtronic ARC catheter measured 20.87 g and 16.78 g respectively. The ACE 068 had highest rate of aspiration at 289 mL/min, and the Catalyst 6 catheter had the lowest rate at 214. The Microvention Sofia catheter had the second highest rate while the ARC had the third highest rate, measuring 285 mL/min and 256 mL/min, respectively.

Conclusions

When using the ADAPT technique, knowledge of the tip force and catheter flow rate of newer catheters with larger distal inner diameters may guide selection of aspiration catheters. While this study demonstrates differences in tip force and flow rate of different commercially available catheters, clinical translation will require further testing and evaluation.

El texto completo de este artículo está disponible en PDF.

Keywords : Thrombectomy, Aspiration, Ischemic stroke, Catheter


Esquema


© 2019  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 49 - N° 3

P. 305-308 - mai 2022 Regresar al número
Artículo precedente Artículo precedente
  • The outcomes of recurrent wide-necked intracranial aneurysms treated with the Woven EndoBridge (WEB): A retrospective bicenter study
  • Kemal Alpay, Alberto Nania, Riitta Parkkola, Jonathan Downer, Antti Lindgren, Riitta Rautio

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

¿Ya suscrito a @@106933@@ revista ?

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.