Accuracy of digitally planned, guided apicoectomy with a conventional trephine and a custom-made endodontic trephine: An in vitro comparative study - 23/08/22
Abstract |
Purpose |
Static guided trephine apicoectomy has been developed as a less invasive and more accurate alternative to conventional freehand apicoectomy with drills. Overpenetration is a frequent issue with this procedure, which deteriorates accuracy and raises safety concerns. A safety improvement to address this problem is presented.
Materials and Methods |
Guided apicoectomies were performed in porcine mandibles with either a conventional bone trephine or a custom-made endo-trephine with built-in depth control. The deviation of the apical endpoint of the trephine from the digital surgical plan was analyzed. Overpenetration frequency was recorded.
Results |
Procedures performed with the custom trephine were significantly more accurate both along the x-axis and globally, but no significant difference was found for the y and z axes. Overpenetration frequency was 70% in the conventional trephine group versus 38% in the stop trephine group.
Conclusion |
The results indicate that the lack of physical depth control can interfere with the accuracy (and safety) of these procedures to a significant extent, as visual cues (such as the depth markings on a conventional trephine) are insufficient to prevent overpenetration. Our results show that custom-made trephines with a built-in stop offer an optimal solution for this problem.
El texto completo de este artículo está disponible en PDF.Keywords : Endodontic microsurgery, Apicoectomy, 3D printing, Static surgical guide