Energy consumption in MRI: Determinants and management options - 24/02/24

Doi : 10.1016/j.neurad.2023.12.001 
Mateusz Chodorowski a, , Julien Ognard a, b, Àlex Rovira c, Jean-Christophe Gentric a, d, David Bourhis d, e, Douraied Ben Salem a, b
a Service d'Imagerie Médicale, CHU Brest, Univ. Brest, Boulevard Tanguy Prigent, Brest CEDEX, 29609, France 
b INSERM UMR 1101, Laboratoire de Traitement de L'Information Médicale - LaTIM, Université de Bretagne Occidentale, 22, Avenue C. Desmoulins, Brest 29238 Cedex 3, France 
c Section of Neuroradiology, Department of Radiology, Hospital Universitari Vall d'Hebron, Universitat Autònoma de Barcelona, Spain 
d Inserm, UMR 1304 (GETBO), Western Brittany Thrombosis Study Group, Université de Bretagne Occidentale, Brest, France 
e Service de Physique Médicale, CHU Brest, France 

Corresponding author.

Abstract

Background

Energy consumption awareness is a known concern, and radiology departments have energy-intensive consuming machines. The means of energy consumption management in MRI scanners have yet to be evaluated.

Purpose

To measure the MRI energy consumption and to evaluate the means to reduce it.

Materials and methods

Data was retrieved for two MRI scanners through the hospital's automated energy consumption measurement software. After correlation with picture archiving and communication system (PACS) files, they were segmented by machine and mode (as follows: stand-by, idle and active) and analyzed. Active mode data for a specific brain MRI protocol have been isolated, and equivalent low energy consuming protocol was made. Both were performed on phantom and compared. Same protocol was performed on a phantom using 3.0T 16 and 32 head channels coils. Multiples sequences were acquired on phantom to evaluate power consumption.

Results

Stand-by mode accounted for 60 % of machine time and 40 % of energy consumption, active mode accounted for 20 % machine time and 40 % energy consumption, idle mode for 20 % imachine time and 20 % consumption. DWI and TOF sequences were the most consuming in our brain-MRI protocol. The low energy consuming protocol allowed a saving of approximately 10 % of energy consumption, which amounted for 0.20€ for each examination. This difference was mainly due to an energy consumption reduction of the DWI sequence. There were no difference in consumption between a 3.0T 16 and 32 channels head coils. Sequence's active power and duration (especially considering slice thickness) have to be taken into account when trying to optimize energy consumption.

Conclusion

There are two key factors to consider when trying to reduce MRI scan energy consumption. Stand-by mode energy consumption has to be taken into account when choosing an MRI scan, as it can't be changed further on. Active mode energy consumption is dependent of the MRI protocols used, and can be reduced with sequences adaptation, which must take into account sequence's active power and duration, on top of image quality.

El texto completo de este artículo está disponible en PDF.

Keywords : MRI, Energy consumption, Green radiology, Stand-by mode

Abbreviations : MRI, DWI, TOF, SWIp, SWAN, PACS, SAR, EPI


Esquema


© 2023  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 51 - N° 2

P. 182-189 - mars 2024 Regresar al número
Artículo precedente Artículo precedente
  • Is non-contrast MRI sufficient to detect meningioma residue after surgery?
  • S. Motillon Alonso, F. Lersy, F.D. Ardellier, H. Cebula, F. Proust, A. Onofrei, A. Chammas, S. Kremer
| Artículo siguiente Artículo siguiente
  • Extended treatment in cerebral ischemia score 2c or 3 as goal of successful endovascular treatment is associated with clinical benefit
  • Adrian Karlsson, Katarina Jood, Isabella M. Björkman-Burtscher, Alexandros Rentzos

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

¿Ya suscrito a @@106933@@ revista ?

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.