Optimisation, validation et application d’une méthode de criblage toxicologique sur DBS par spectrométrie de masse haute résolution (Orbitrap Exploris 120) - 14/11/26

Optimisation, validation and application of a toxicological screening method on dried blood spots by high-resolution mass spectrometry (Orbitrap Exploris 120)

Doi : 10.1016/j.toxac.2026.11.003 
Doudka Natalia 1, ⁎ , Pelissier-Alicot Anne-Laure 2, Bayo Gnominding 1, Solas-Chesneau Caroline 1, Fabresse Nicolas 1, 3
1 Laboratoire de Pharmacocinétique et de Toxicologie, CHU La Timone, 264 rue Saint-Pierre, 13005, Marseille, France 
2 Service de Médecine Légale, CHU La Timone, 264 rue Saint-Pierre, 13005, Marseille, France 
3 Aix Marseille Université, INSERM, IRD, SESSTIM, Sciences Economiques et Sociales de la Santé & Traitement de l’Information Médicale, Marseille, France 

⁎ Auteur correspondant: CHU La Timone, Laboratoire de pharmacocinétique et de toxicologie, 264 rue Saint-Pierre, 13005 Marseille, France CHU La Timone, Laboratoire de pharmacocinétique et de toxicologie, 264 rue Saint-Pierre Marseille 13005 France
En prensa. Manuscrito Aceptado. Disponible en línea desde el Saturday 14 November 2026

Résumé

Objectif : Développer et valider une méthode de criblage toxicologique sur gouttes de sang séché (Dried Blood Spot,DBS) par chromatographie liquide couplée à la spectrométrie de masse haute résolution, et la comparer sur sang post-mortem à une méthode utilisée en routine (extraction QuEChERS).

Méthodes : Dix microlitres de sang déposés sur papier Whatman 903 ont été séchés, découpés en totalité, extraits au méthanol sous ultrasons (30 min) et analysés en mode data-dependent. La méthode a été validée selon l’ICH M10 (International Council for Harmonisation Multidisciplinary Guideline) pour 20 molécules, puis appliquée à 114 échantillons de sang post-mortem .

Résultats : La linéarité était satisfaisante (r² > 0,99) sur des gammes de 0,5-50, 5-500 ou 50-5 000 ng/mL selon les molécules. Les coefficients de variation et biais intra- et inter-séries étaient inférieurs à 15 %, les rendements de 10 à 83 % et les facteurs matrice normalisés de 0,8 à 8,2. Les composés étaient stables au moins 48 h à température ambiante, −20 °C et −80 °C. Le criblage a identifié 135 des 144 molécules testées (limite d’identification médiane 10 ng/mL). Sur les sangs post-mortem , la concordance qualitative était de 98,1 % (317 identifications, aucun faux positif). Sur 64 dosages appariés, la corrélation était étroite (r = 0,98), mais les concentrations sur DBS étaient supérieures d’environ 30 % (pente de Passing-Bablok 1,32 ; IC 95 % 1,22-1,43), surtout pour les opioïdes.

Conclusion : Simple, peu invasive et économique, cette méthode est un outil prometteur pour le criblage toxicologique de routine, sous réserve de prendre en compte ce biais pour toute interprétation quantitative.

El texto completo de este artículo está disponible en PDF.

Abstract

Aim: To develop and validate a toxicological screening method on dried blood spots (DBS) using liquid chromatography coupled to high-resolution mass spectrometry, and to compare it on post-mortem blood with a routine method (QuEChERS extraction).

Methods: Ten microlitres of blood spotted onto Whatman 903 paper were dried, fully punched, extracted with methanol under sonication (30 min) and analyzed in data-dependent mode. The method was validated according to ICH M10 for 20 compounds, then applied to 114 post-mortem blood samples.

Results: Linearity was satisfactory (r² > 0.99) over ranges of 0.5-50, 5-500 or 50-5000 ng/mL depending on the compound. Intra- and inter-assay coefficients of variation and bias were below 15 %, recoveries ranged from 10 to 83 % and normalized matrix factors from 0.8 to 8.2. Compounds were stable for at least 48 h at room temperature, −20 °C and −80 °C. Screening identified 135 of 144 tested compounds (median limit of identification 10 ng/mL). In the post-mortem samples, qualitative agreement was 98.1 % (317 identifications, no false positive). Across 64 paired results, correlation was close (r = 0.98), but DBS concentrations were about 30 % higher (Passing-Bablok slope 1.32; 95 % CI 1.22-1.43), mainly for opioids.

Conclusion: Simple, minimally invasive and cost-effective, this method is a promising tool for routine toxicological screening, provided that this bias is taken into account for any quantitative interpretation.

El texto completo de este artículo está disponible en PDF.

Mots-clés : Goutte de sang séché, Criblage toxicologique, Spectrométrie de masse haute résolution, Toxicologie médico-légale, Sang post-mortem

Keywords : Dried blood spot testing, Toxicological screening, High-resolution mass spectrometry, Forensic toxicology, Post-mortem blood



© 2026  Société Française de Toxicologie Analytique. Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

¿Ya suscrito a @@106933@@ revista ?

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.