Suscribirse

Metal-on-metal bearings total hip arthroplasty: The cobalt and chromium ions release concern - 30/11/10

Doi : 10.1016/j.otsr.2010.05.008 
C. Delaunay a, , I. Petit b, I.D. Learmonth c, P. Oger d, P.A. Vendittoli e
a De l’Yvette Private Hospital, 67, route de Corbeil, 91160 Longjumeau, France 
b Biomnis Biological Laboratory, 78, avenue de Verdun, 94200 Ivry-sur-Seine, France 
c Bristol Implant Research Centre, Avon Orthopaedic Centre, Southmead Hospital, Westbury-on-Trym, Bristol, BS10 5NB, United Kingdom 
d Department of Orthopaedic Surgery, Versailles Hospital, 177, rue de Versailles, 78150 Le Chesnay, France 
e Orthopaedics Division, Montréal University, Maisonneuve-Rosemont Hospital, 5415 boulevard de L’Assomption, Montréal, Québec H1T 2M4, Canada 

Corresponding author. Tel.: +08 26 30 60 30; fax: +01 69 10 30 20.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Summary

With certain concerns recently reported on metal-on-metal bearing couples in total hip arthroplasty, this study’s objective is to review the current knowledge concerning release of metal ions and its potential consequences. Each metal-on-metal implant presents different tribological properties. The analytical techniques for metals are accurate and the Co ion rates seem acceptable up to 2μg/L. A delayed type IV hypersensitivity reaction (atypical lymphocytic vasculitis-associated lesion [ALVAL]) may be the source of arthroplasty failure. Idiosyncratic, it remains unpredictable even using cutaneous tests and apparently is rare (0.3%). Today, there are no scientific or epidemiologic data supporting a risk of carcinogenesis or teratogenesis related to the use of a metal-on-metal bearings couple. Solid pseudotumors nearly exclusively are observed with resurfacing procedures, carrying a high annual revision rate in women under 40 years of age, occurring particularly in cases of acetabular malposition and with use of cast molded Cr-Co alloys. Osteolysis manifests through complete and progressive radiolucent lines or through cavitary lesions stemming from ALVAL-type alterations or impingement problems or implant incompatibility. The formation of wear debris exceeding the biological tolerance is possible with implant malposition, subluxation, and jamming of the femoral head in cases of cup deformity. Moreover, each implant presents different metal ion production; assessment of their performance and safety is required before their clinical use. With the knowledge available today, metal-on-metal bearing couples are contraindicated in cases of metal allergies or end stage renal dysfunction and small size resurfacing should cautiously be used.

El texto completo de este artículo está disponible en PDF.

Keywords : Total hip arthroplasty, Metal-on-metal bearings, Hypersensitivity, Pseudotumors, Resurfacing


Esquema


 Presented at the “Technologic Watch” session of the French Hip and Knee Society (Société française de la hanche et du genou [SFHG]), during the Speciality day of the SoFCOT annual convention, on November, 2009.


© 2010  Publicado por Elsevier Masson SAS.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 96 - N° 8

P. 894-904 - décembre 2010 Regresar al número
Artículo precedente Artículo precedente
  • Deep venous thrombosis associated with acute hematogenous osteomyelitis in children
  • S. Bouchoucha, F. Benghachame, M. Trifa, W. Saied, W. Douira, M.N. Nessib, M.B. Ghachem
| Artículo siguiente Artículo siguiente
  • Giant cells tumor recurrence at the third lumbar vertebra
  • M. Mestiri, M. Bouabdellah, R. Bouzidi, S. Kammoun, N. Mnif, S. Hawet, M. Kooli, M. Zlitni

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.