Arthroscopic dynamic analysis of scapular notching in reverse shoulder arthroplasty - 08/12/11

Doi : 10.1016/j.otsr.2011.07.014 
O. Boughebri a, , F. Duparc b, J.-M. Adam b, P. Valenti c
a Armand Brillard Private Hospital, Orthopaedic Surgery and Sports Medicine Department, 3, avenue Watteau, 94130 Nogent-sur-Marne, France 
b Orthopaedic Surgery and Traumatology Department, Charles-Nicolle Teaching Hospital Center, 1, rue de Germont, 76000 Rouen, France 
c Hand Surgery Institute, Jouvenet Private Hospital, 6, Square Jouvenet, 75016 Paris, France 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Summary

Introduction

Despite the success of reverse shoulder arthroplasty, the rate of scapular notching phenomena is high and represents a significant risk of glenoid prosthetic component loosening over the long term. The aim of this study was to perform an arthroscopic dynamic analysis of the impingement of the humeral cup on the neck of the scapula in order to highlight the causes of the occurrence of this notch and possible ways to avoid it.

Materials and methods

The SMR reverse shoulder prosthesis (Lima®) was successively implanted in 11 shoulders of seven cadavers. Three glenospheres (36mm, 36mm with a 4-mm lowering offset, and 44mm) were consecutively tested on each shoulder. An arthroscopic study was then performed in each case looking for scapular notching depending on the position of the shoulder, its exact location, and any resulting prosthetic instability. Rates of notching for each glenosphere in the different shoulder positions were compared.

Results

For each glenosphere, the highest rates of notching were recorded in the zero abduction position. Increasing abduction reduced the rates of notching and prosthetic instability for all types of glenosphere. External rotation increased both risks with 36-mm glenospheres but reduced them with the 44-mm glenosphere. The lower center of rotation, thus, reduced the risk of notching but increased the risk of instability associated with it. Increasing the prosthetic diameter reduced both risks.

Conclusion

Reducing the risk of scapular notching and prosthetic instability requires, in addition to a lower center of rotation, the use of glenosphere diameters greater than 36mm. The use of a humeral cup with a posteroinferior indentation should also contribute to reducing this risk.

Level of evidence

Level IV.

El texto completo de este artículo está disponible en PDF.

Keywords : Reverse shoulder arthroplasty, Scapular notching, Shoulder arthroplasty instability


Esquema


© 2011  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 97 - N° 8

P. 779-784 - décembre 2011 Regresar al número
Artículo precedente Artículo precedente
  • Contents
| Artículo siguiente Artículo siguiente
  • Free Gracilis muscle transfer to restore elbow flexion in brachial plexus injuries
  • B. Coulet, C. Boch, J. Boretto, C. Lazerges, M. Chammas

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.