Diffusion-weighted MR imaging of the normal pancreas: Reproducibility and variations of apparent diffusion coefficient measurement at 1.5- and 3.0-Tesla - 30/03/13

Doi : 10.1016/j.diii.2012.12.007 
M. Barral a, P. Soyer b, , c, d , W. Ben Hassen a, E. Gayat c, e, M. Aout f, M. Chiaradia a, g, h, A. Rahmouni a, g, h, A. Luciani a, g, h
a Department of Radiology, Groupe Henri-Mondor Albert-Chenevier, AP–HP, 51, avenue du Maréchal-de-Lattre-de-Tassigny, 94010 Créteil cedex, France 
b Department of Abdominal Imaging, Hôpital Lariboisière, AP–HP, 2, rue Ambroise-Paré, 75010 Paris, France 
c Université Paris-Diderot, Sorbonne Paris Cité, 10, avenue de Verdun, 75010 Paris, France 
d UMR Inserm 965-Paris 7 « Angiogenèse et recherche translationnelle », 2, rue Amboise-Paré, 75010 Paris, France 
e Department of Anesthesiology and Intensive Care Medicine, Hôpital Lariboisière, AP–HP, 2, rue Ambroise-Paré, 75010 Paris, France 
f Clinical Research Unit, Hôpital Lariboisière, AP–HP, 2, rue Ambroise-Paré, 75010 Paris, France 
g Faculté de médecine, université Paris Est Créteil, 94010 Créteil, France 
h Inserm unité U 955, équipe 17, 94010 Créteil, France 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Purpose

To evaluate reproducibility and variations in apparent diffusion coefficient (ADC) measurement in normal pancreatic parenchyma at 1.5- and 3.0-Tesla and determine if differences may exist between the four pancreatic segments.

Materials and methods

Diffusion-weighted MR imaging of the pancreas was performed at 1.5-Tesla in 20 patients and at 3.0-Tesla in other 20 patients strictly matched for gender and age using the same b values (0, 400 and 800s/mm2). Two independent observers placed regions of interest within the four pancreatic segments to measure ADC at both fields. Intra- and inter-observer agreement in ADC measurement was assessed using Bland-Altman analysis and comparison between ADC values obtained at both fields using non-parametrical tests.

Results

There were no significant differences in ADC between repeated measurements and between ADC obtained at 1.5-Tesla and those at 3.0-Tesla. The 95% limits of intra-observer agreement between ADC were 2.3%–22.7% at 1.5-Tesla and 1%–24.2% at 3.0-Tesla and those for inter-observer agreement between 1.9%–14% at 1.5-Tesla and 8%–25% at 3.0-Tesla. ADC values were similar in all pancreatic segments at 3.0-T whereas the tail had lower ADC at 1.5-Tesla.

Conclusion

ADC measurement conveys high degrees of intra- and inter-observer reproducibility. ADC have homogeneous distribution among the four pancreatic segments at 3.0-Tesla.

El texto completo de este artículo está disponible en PDF.

Keywords : Diffusion-weighted MRI, MR imaging, Apparent diffusion coefficient, Pancreas, Reproducibility assessment


Esquema


© 2012  Éditions françaises de radiologie. Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 94 - N° 4

P. 418-427 - avril 2013 Regresar al número
Artículo precedente Artículo precedente
  • Evaluation of the value of abdominopelvic acquisition without contrast injection when performing a whole body CT scan in a patient who may have multiple trauma
  • P. Naulet, J. Wassel, A. Gervaise, A. Blum
| Artículo siguiente Artículo siguiente
  • Wormian bones in a general paediatric population
  • B. Marti, D. Sirinelli, L. Maurin, E. Carpentier

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.