Accuracy of continuous subcutaneous glucose monitoring with the GlucoDay® in type 1 diabetic patients treated by subcutaneous insulin infusion during exercise of low versus high intensity - 17/02/08

Doi : DM-09-2006-32-4-1262-3636-101019-200518933 

C Fayolle [1 et 2],

JF Brun [1],

J Bringer [2],

J Mercier [1],

E Renard [2]

Ver las filiaciones

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Exactitude de la surveillance continue du glucose sous-cutané par le GlucoDay® chez des patients diabétiques de type 1 traités par perfusion sous-cutanée d'insuline durant un exercice de faible ou haute intensité

Objectif

Le GlucoDay permet une surveillance continue du glucose par microdialyse sous-cutanée dans des conditions sédentaires. Pour le valider lorsque la glycémie est exposée à des variations rapides et importantes, nous avons exploré son exactitude durant deux séances d'exercice physique avec utilisation du glucose très différente.

Methodes

Neuf patients diabétiques masculins, âgés de 32 à 61 ans, traités par pompe à insuline, ont d'abord réalisé un exercice physique maximal standard destiné à déterminer la consommation maximale d'oxygène et le premier seuil ventilatoire (Vt1). Ensuite deux épreuves d'exercice de 30 minutes en plateau à 15 % en dessous et 15 % au dessus du Vt1 ont été réalisées en ordre aléatoire avec le GlucoDay, et mesure des taux d'oxydation glucidique par calorimétrie indirecte.

Resultats

L'oxydation glucidique durant l'exercice à +15 % Vt1 a été plus importante (+943.5 mg/min, soit +45.5 %, P ≪ 0.01) que durant l'exercice à -15 % Vt1. Aucune hypoglycémie n'est survenue. Suite à la fracture de 39 % des sondes sous-cutanées, onze séances d'exercice en plateau chez 7 sujets ont permis de comparer 141 valeurs appariées de la glycémie (d'après le sensor et le sang veineux). La grille d'erreur de Clarke situe 92,9 % des valeurs de glucose appariées dans la zone A et 6,4 % des valeurs dans la zone B, tandis qu'un seul couple de valeurs (0,7 %) est dans la zone D. La glycémie veineuse tendait à décroître plus rapidement que la glycémie d'après le sensor durant l'exercice. L'analyse selon Bland-Altman montrait quelques cas de sous-estimation de la glycémie veineuse lors de l'exercice de haute intensité.

Conclusions

Cette étude montre lors de l'exercice une exactitude satisfaisante du GlucoDay. Un discret délai de réponse dans les valeurs du sensor explique probablement quelques différences qui n'apparaissent pas cliniquement significatives. Une moindre fragilité des sondes et la confirmation de l'exactitude en hypoglycémie soutiendraient plus avant l'utilisation du GlucoDay® lors de l'exercice.

Aim

The GlucoDay allows continuous glucose monitoring by subcutaneous microdialysis in sedentary conditions. To validate it when glycaemia may undergo rapid and dramatic changes, we investigated its accuracy during two exercise sessions with markedly different glucose disposal rates.

Methods

Nine male diabetic patients, aged 32-61, treated by insulin pumps, first underwent a standard maximal exercise-test designed for determining the maximal oxygen consumption and the first ventilatory threshold (Vt1). Then two 30 min steady-state workloads at 15% below and 15% above the Vt1 were performed in random order with the GlucoDay, and measurement of CHO oxidation rates was made by indirect calorimetry.

Results

CHO oxidation during exercise at +15% Vt1 was higher (+943.5 mg/min, ie +45.5%, P≪0.01) than during exercise at -15% Vt1 No hypoglycaemia occurred. Due to breakages of 39% of subcutaneous probes, eleven steady-state sessions in 7 subjects allowed to compare 141 paired glucose (sensor vs. venous) determinations. The Clarke error grid situates 92.9% of glucose values within the A zone and 6.4% within the B zone, while only one pair of values (0.7%) falls in the D zone. Venous glucose tended to decrease more rapidly than sensor glucose during exercise. Bland-Altman plots evidence for a few cases of underestimation of venous glucose at high intensity.

Conclusions

This study showed satisfactory accuracy of the GlucoDay during exercise. A slight lag time in sensor values likely explains a few discrepancies that do not appear as clinically meaningful. Reduction of probe fragility and confirmed sensor accuracy in hypoglycaemia would further support applicability of GlucoDay at exercise.


Mots clés : Détecteur de glucose , Microdialyse , Exercice physique , Pompe à insuline , Calorimétrie indirecte

Keywords: Glucose sensor , Microdialysis , Exercise , Insulin pump , Indirect calorimetry


Esquema



© 2006 Elsevier Masson SAS. Tous droits réservés.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 32 - N° 4

P. 313-320 - septembre 2006 Regresar al número
Artículo precedente Artículo precedente
  • The refusal of treatment in anorexia nervosa, an ethical conflict with three characters: “the girl, the family and the medical profession”.
  • B Vialettes, C Samuelian-Massat, R Valéro, S Béliard
| Artículo siguiente Artículo siguiente
  • Iodine status, thyroid function, thyroid volume and thyroid autoimmunity in patients with type 1 diabetes mellitus in an iodine-replete area
  • A Okten, S Akcay, M Cakir, I Girisken, P Kosucu, O Deger

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.