Achieving Targeted Pulse Oximetry Values in Preterm Infants in the Delivery Room - 23/07/13

Doi : 10.1016/j.jpeds.2013.01.010 
Bheru Gandhi, MD, Wade Rich, RRT , Neil Finer, MD
Division of Neonatology, University of California San Diego Medical Center, San Diego, CA 

Reprint requests: Wade Rich, RRT, CCRC, Division of Neonatology, UCSD Medical Center, 402 Dickinson Street, MPF 1-140, #8774, San Diego, CA 92103.

Abstract

Objective

To determine whether resuscitation teams can better maintain newborn transitional oxygen saturation (SpO2) values within a prespecified target range using a graphical display of the targets and real-time SpO2 data compared with using only numerical oximeter values.

Study design

Preterm neonates were enrolled in this prospective cohort evaluation of a change in practice. The Transitional Oxygen Targeting System (TOTS) plots real-time SpO2 values in relation to 10th and 50th percentile SpO2 curves, which provides a visual target. After introduction of the TOTS, the resuscitation team adjusted the fraction of inspired oxygen to maintain the SpO2 within the target range, and before its use neonates were resuscitated in attempt to target normal transitional SpO2 values without using the TOTS monitor. Duration of time within 10th-50th percentile values was compared between 2 cohorts, children evaluated with the TOTS and those not evaluated with the TOTS (controls).

Results

A total of 40 infants were enrolled, including 20 prospectively evaluated with the TOTS and 20 controls. Mean gestational age, birth weight, total resuscitation time, duration of supplemental oxygen administration, changes in oxygen concentration, and respiratory support provided were similar in the 2 groups. The TOTS cohort spent 52% of time within the target range; the control cohort, 37% (P = .03).

Conclusion

SpO2 values were maintained within a specified target range for significantly longer in preterm neonates resuscitated using the TOTS display compared with those resuscitated without TOTS.

El texto completo de este artículo está disponible en PDF.

Keyword : FiO2, GA, NRP, ROAR, SpO2, TOTS


Esquema


 The authors declare no conflicts of interest.


© 2013  Mosby, Inc. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 163 - N° 2

P. 412-415 - août 2013 Regresar al número
Artículo precedente Artículo precedente
  • Enteral Feeding during Indomethacin and Ibuprofen Treatment of a Patent Ductus Arteriosus
  • Ronald Clyman, Andrea Wickremasinghe, Nami Jhaveri, Denise C. Hassinger, Joshua T. Attridge, Ulana Sanocka, Richard Polin, Maria Gillam-Krakauer, Jeff Reese, Mark Mammel, Robert Couser, Neil Mulrooney, Toby D. Yanowitz, Matthew Derrick, Priya Jegatheesan, Michele Walsh, Alan Fujii, Nicolas Porta, William A. Carey, Jonathan R. Swanson, Ductus Arteriosus Feed or Fast with Indomethacin or Ibuprofen (DAFFII) Investigators ?
| Artículo siguiente Artículo siguiente
  • Effects of Different Complementary Feeding Regimens on Iron Status and Enteric Microbiota in Breastfed Infants
  • Nancy F. Krebs, Laurie G. Sherlock, Jamie Westcott, Diana Culbertson, K. Michael Hambidge, Leah M. Feazel, Charles E. Robertson, Daniel N. Frank

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

¿Ya suscrito a @@106933@@ revista ?

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.