Suscribirse

Medium-term osteolysis with the Wallaby I® deep-dished total knee prosthesis - 30/04/14

Doi : 10.1016/j.otsr.2014.03.014 
B. Marion a, b, D. Huten c, P. Boyer a, b, C. Jeanrot a, b, P. Massin a, , b
a Service de chirurgie orthopédique, hôpital Bichat-Claude-Bernard, 46, rue Henri-Huchard, 75877 Paris cedex 18, France 
b EA REMES, université Paris-Diderot, Sorbonne Paris Cité, 75010 Paris, France 
c Service de chirurgie orthopédique, CHU de Rennes, 16, boulevard Bulgarie, BP 90347, 35203 Rennes cedex 2, France 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

En prensa. Pruebas corregidas por el autor. Disponible en línea desde el Wednesday 30 April 2014
This article has been published in an issue click here to access

Abstract

Background

Highly congruent total knee prostheses were introduced in the 1990s in the hope of decreasing polyethylene wear, thereby minimising loosening and particle-induced peri-prosthetic osteolysis. Despite promising long-term outcomes, substantial rates of aseptic loosening were reported with conventional gamma-irradiated polyethylene inserts, suggesting that highly reticulated polyethylene should be used instead. We assessed medium-term outcomes of the Wallaby I® total knee prosthesis with a deep-dished tibial insert made of conventional gamma-irradiated polyethylene.

Hypothesis

We hypothesised that the deep-dished Wallaby I® prosthesis was associated with similar or lower rates of aseptic loosening and peri-prosthetic osteolysis compared to posterior-stabilised prostheses.

Materials and methods

At our institution, 121 consecutive patients underwent total knee arthroplasty (TKA) with a deep-dished cemented prosthesis (Wallaby I®, Sulzer/Centerpulse, Zürich, Switzerland) between 2001 and 2005. Among them, 89 had complete follow-up data over a 4-year period and a mean follow-up of 96months. We retrospectively analysed the clinical and radiographic IKS scores in these 89 patients.

Results

Osteolysis with aseptic loosening required revision TKA of 10 knees after a mean follow-up of 81months. Mean 9-year prosthesis survival was 88±17%. Four inserts exhibited evidence of delamination. A fracture of the postero-medial aspect of the tibial baseplate beneath a zone of insert wear was found in 1 knee and gross mobility of the insert on the baseplate in 6 knees. The other 79 patients had good clinical and radiographic outcomes with a mean range of active knee flexion of 108±15°.

Discussion

The medium-term outcomes in our study were inferior to those reported with posterior-stabilised tibial components. Sporadic variations in polyethylene quality may explain the cases of osteolysis (shelf oxidation). In addition, the increased shear stresses related to the deep-dish design may increase backside wear, thereby compromising insert fixation to the baseplate. We believe the Wallaby I® prosthesis should no longer be used, and we recommend computed tomography follow-up of patients harbouring this prosthesis.

Level of evidence

Level IV (retrospective study).

El texto completo de este artículo está disponible en PDF.

Keywords : Deep-dished total knee arthroplasty, Particle-induced osteolysis, Loosening, Revision total knee arthroplasty


Esquema


© 2014  Publicado por Elsevier Masson SAS.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.