Cruciate-sacrificing total knee arthroplasty and insert design: A radiologic study of sagittal laxity - 08/12/15

Doi : 10.1016/j.otsr.2015.07.024 
B. Appy Fedida a, b, , E. Krief c, E. Havet c, P. Massin a, b, P. Mertl c
a Service de chirurgie orthopédique, hôpital Bichat-Claude Bernard, 46, rue Henri-Huchard, 75877 Paris cedex 18, France 
b EA REMES, université Paris-Diderot, Sorbonne Paris Cité, 75010 Paris, France 
c Service de chirurgie orthopédique, CHU d’Amiens, avenue René-Laënnec-Salouel, 80054 Amiens, France 

Corresponding author. Service de chirurgie orthopédique, hôpital Bichat-Claude Bernard, 46, rue Henri-Huchard, 75877 Paris cedex 18, France.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Introduction

Ultracongruent inserts avoid some of the drawbacks of central spine postero-stabilized inserts. However, early wear has been reported, and may be due to increased sagittal laxity. The principal objective of the present study was to compare sagittal laxity in rotating platform total knee replacements (TKR) according to insert design: ultracongruent versus central spine. The principal hypothesis was that insert design influences global sagittal laxity.

Material and methods

A retrospective comparative study recruited 3 consecutive series of patients treated for primary osteoarthritis of the knee, with a minimum 1 year's follow-up. The UC series comprised 35 knees in 34 patients, receiving a Total Knee Triathlon™ (Stryker Orthopaedics, Mahwah, NJ) TKR with ultracongruent insert, at a mean 2.0 years’ follow-up. The UC+ series comprised 36 knees in 34 patients, receiving the BalanSys™ (Mathys Ltd, Bettlach, Switzerland) TKR with ultracongruent insert, at a mean 2.5 years’ follow-up; in this model, the anterior edge of the insert is higher than in the UC series (“deep-dish” design). The PS series comprised 43 knees in 40 patients, receiving a Total Knee Triathlon™ (Stryker Orthopaedics, Mahwah, NJ) TKR with central spine posterior stabilization, at a mean 1.5 years’ follow-up. The principal assessment criterion was sagittal laxity at 90° flexion as measured by the Telos Stress Device® (Metax GmbH, Hungen, Germany).

Results

Sagittal laxity did not significantly differ between the UC and UC+ series: mean 8.2mm (range: 0–19.5mm) and 8.4mm (4.5–15.8mm), respectively. Sagittal laxity in the PS series was significantly less: 1.4mm (0.2–3.9) (P<0.0001).

Conclusion

Sagittal laxity was greater in ultracongruent than central spine posterior stabilized TKR. This anteroposterior movement may induce polyethylene wear. The ideal degree of sagittal laxity for ultracongruent inserts remains to be determined.

Level of evidence

IV – retrospective study.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Total knee replacement, Posterior stabilization, Ultracongruent insert, Design, Sagittal laxity


Mappa


© 2015  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 101 - N° 8

P. 941-945 - dicembre 2015 Ritorno al numero
Articolo precedente Articolo precedente
  • Poor reproducibility of the MRI measurement of distal femoral torsion
  • M. Ollivier, C. Stelzlen, P. Boisrenoult, N. Pujol, P. Beaufils
| Articolo seguente Articolo seguente
  • Does primary total knee arthroplasty for acute knee joint fracture maintain autonomy in the elderly? A retrospective study of 21 cases
  • F. Boureau, K. Benad, S. Putman, G. Dereudre, G. Kern, C. Chantelot

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2025 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.