Pull-out strength of four tibial fixation devices used in anterior cruciate ligament reconstruction - 31/03/18

Doi : 10.1016/j.otsr.2017.12.011 
M. Chivot a, b, S. Harrosch a, b, F. Kelberine c, M. Pithioux a, b, J.-N. Argenson a, b, M. Ollivier a, b,
a Aix-Marseille université, CNRS, ISM UMR 7287, 13288 Marseille cedex 09, France 
b Service de chirurgie orthopédique et traumatologique, institut du mouvement et de la locomotion, hôpital Sainte-Marguerite, 270, boulevard Sainte-Marguerite, BP 29, 13274 Marseille, France 
c Clinique La-Provençale, 67, Cours-Gambetta, 13100 Aix-en-Provence, France 

Corresponding author.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Introduction

In reconstructions of the anterior cruciate ligament (ACL), tibial fixation can be the weak point in the assembly during the early postoperative period. The present study sought to compare pull-out strength between four tibial fixation systems used in ACL reconstruction.

Hypothesis

The study hypothesis was that all four devices show ≥450N pull-out strength with comparable biomechanical breakage characteristics.

Material and methods

An experimental study used a mechanical model to perform axial traction on a synthetic ligament (polypropylene cord folded in four) implanted in an artificial tibia (Sawbones Proximal Tibia # 1116-2: model: normal anatomy; solid foam; size: medium) using four tibial fixation systems: Ligafix® interference screw (SBM™); Bio-Intrafix® (Mitek™); Translig® (SBM™); RIGIDfix® (SBM™). For each system, four models were tested using an Instron 5566® traction machine, allowing 100mm/min stretching up to breakage. Study parameters comprised: pull-out strength, maximal whole assembly slippage, stiffness at breaking point, and type of break.

Results

Mean pull-out strength was 450±24N (range, 421–488N) for Ligafix®, 415±60N (327–454N) for Bio-Intrafix®, 539±66N (449–636N) for RigidFix and 1067±211N (736–1301N) for Translig®, and was significantly greater for Translig® than for the other devices (p=0.02), which did not significantly differ from one another. The expected maximal load of 450N was reached in 100% of cases with Translig® and RIGIDfix® and in 50% of cases with Bio-Intrafix® and Ligafix®. There were no significant differences regarding stiffness. Ligafix® showed significantly less slippage than the others (p=0.006), with breakage caused by the ligament sliding between bone and implant.

Discussion

In this in-vitro study, the Translig® fixation device showed better pull-out strength than the other three devices tested.

Type of study and level of evidence

Comparative laboratory study. Level II.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Biomechanics, Fixation, Tibia, Anterior cruciate ligament reconstruction, Pull-out strength


Mappa


© 2018  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 104 - N° 2

P. 203-207 - aprile 2018 Ritorno al numero
Articolo precedente Articolo precedente
  • Complications and revision surgeries in two extra-articular tenodesis techniques associated to anterior cruciate ligament reconstruction. A case-control study
  • C. Batailler, S. Lustig, O. Reynaud, P. Neyret, E. Servien
| Articolo seguente Articolo seguente
  • Anterior cruciate ligament revision with Achilles tendon allograft in young athletes
  • S. Zaffagnini, A. Grassi, G.M. Marcheggiani Muccioli, T. Roberti Di Sarsina, L. Macchiarola, M. Mosca, M.P. Neri, M. Marcacci

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2025 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.