Abbonarsi

InSafeLock humeral nail provides a safe application for proximal and distal locking screws with distal endopin – An anatomical study - 27/08/19

Doi : 10.1016/j.otsr.2019.04.014 
Murat Tasci a, , İsmail Turkmen a, Haluk Celik a, Mehmet Akif Akcal b, Rahime Şekerci c, Nigar Keles c, Necdet Saglam a, Fuat Akpinar a
a Umraniye Training and Research Hospital, Orthopaedics and Traumatology, Istanbul, Turkey 
b Antalya Ataturk State Hospital, Orthopaedics and Traumatology, Antalya, Turkey 
c Akdeniz University, Anatomy, Antalya, Turkey 

Corresponding author. Atatürk mah. Guner sokak, Metropark Sitesi A2 Blok D10, Kuçukçekmece, Istanbul, Turkey.Atatürk mah. Guner sokakMetropark Sitesi A2 Blok D10KuçukçekmeceIstanbulTurkey

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Introduction

Efforts to prevent iatrogenic neurovascular injuries with humeral intramedullary nailing lead to design new implants and inside to out distal locking technique using an endopin aims to provide a safer screw application. InSafeLock (TST, Istanbul, Turkey) humeral nail have been recently developed to minimize the possible screw related complications. The anatomical relationship between locking screws and neurovascular structures with the application time were compared between Trigen Humeral Nail (Smith and Nephew, Memphis, USA) and InSafeLock Humeral Nail.

Hypothesis

InSafeLock humeral nail would be safer than Trigen Humeral nail in terms of neurovascular injury.

Materials and methods

Seven cadavers were used with both shoulders and surgical application of two nails was performed as the manufacturer guide. An Insafelock humeral nail was used for each right humerus and a Trigen humeral nail was used for each left humerus. Once the nails were placed, proximal and distal region of the nails were dissected to evaluate the relationship between screws and adjacent anatomical structures. The duration of the each screw was assessed via a stopwatch.

Results

No significant finding was noted for the relationship between the neurovascular structures and proximal screws in two groups (p<0.05). The distal locking of the InSafeLock humerus nail had a shorter application time and no neurovascular damage was recorded.

Discussion

The newly developed Insafelock humerus nails are at least as safe and effective as current humeral nails available on the market. Additional benefits include the preservation of neurovascular structures, as the Insafelock humerus nail does not require the use of an extra incision. Furthermore, surgical time is significantly shorter with using distal endopin.

Level of study

III, controlled laboratory study.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Humerus diaphysis fracture, Intramedullary nail, Neurovascular damage


Mappa


© 2019  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 105 - N° 5

P. 1005-1011 - settembre 2019 Ritorno al numero
Articolo precedente Articolo precedente
  • Femoral neck shortening after internal fixation of Garden I fractures increases the risk of femoral head collapse
  • Léo Nanty, François Canovas, Thibaut Rodriguez, Patrick Faure, Louis Dagneaux
| Articolo seguente Articolo seguente
  • Does migration of osseointegrated implants for transfemoral amputees predict later revision? A prospective 2-year radiostereometric analysis with 5-years clinical follow-up
  • Rehne Lessmann Hansen, Bente Lomholt Langdahl, Peter Holmberg Jørgensen, Klaus Kjær Petersen, Kjeld Søballe, Maiken Stilling

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2025 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.