Abstract

Introduction

To avoid the DRUJ surgical procedures disadvantages, a new intraosseous distal radioulnar prosthesis designed on Sauvé-Kapandji procedure has been introduced. Stability of the prosthesis and biomechanics are to be evaluated in this article.

Materials and methods

On a cadaveric study, during placement of the prosthesis, biometry of the bones, prosthesis stability (in axial and lateral tractions, wrist pronation and supination, and squeeze test), wrist range of motion before and after implantation, and radiographic evaluation were done on 16 cadavers.

Results

Range of motion of the wrist joint before and after the insertion of the prosthesis, had no significant difference in all six directions. Stability of the prosthesis, when rotational pronation force was exerted, was greater than when rotational supination force was exerted. The prosthesis showed significant stability against longitudinal traction forces in a way that no prosthesis dislocation was observed up to 150N forces. Stability of the prosthesis was investigated when lateral force was applied to different wrist positions. The most stable position of the prosthesis was in the case of lateral traction forces in supination where no case of dislocation was observed.

Conclusions

The intraosseous distal radioulnar prosthesis demonstrated stable structure with no effect on wrist range of motion.

Level of evidence

IV.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Distal radioulnar prosthesis, Cadaver study, Kapandji procedure


Mappa


© 2021  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 107 - N° 5

Articolo 102975- settembre 2021 Ritorno al numero
Articolo precedente Articolo precedente
  • Four-corner fusion: Clinical and radiological outcome after fixation by headless compression screws or dorsal locking plate at minimum 5 years’ follow-up
  • Marc-Antoine d’Almeida, Nadine Sturbois-Nachef, Thomas Amouyel, Christophe Chantelot, Marc Saab
| Articolo seguente Articolo seguente
  • Comparison of subjective outcomes of Darrach and Sauvé–Kapandji procedures at a minimum 2 years’ follow-up
  • Steven Roulet, Ann Williot, Clara Sos, Matthieu Mazaleyrat, Emilie Marteau, Jacky Laulan, Guillaume Bacle

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2025 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.