A novel method to improve femoral head and stem taper stability intraoperatively in total hip arthroplasty – a proof of concept study - 25/11/22

Doi : 10.1016/j.otsr.2022.103284 
Shouchen Dun , Boon Him Lim, Stephen W. Swope, Dustin R. Whitaker
 DePuy Synthes Joint Reconstruction, 700 Orthopaedic Drive, Warsaw, IN 46581, USA 

Corresponding author.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Background

Mechanically assisted crevice corrosion (MACC) has been associated with the compromised durability and fixation of modular total hip implants, adverse reaction of local tissue, and other undesirable clinical outcomes in total hip arthroplasty (THA). MACC is primarily caused by the relative motion between the femoral head and stem. To minimize the relative motion the taper connection between the two components must be strong enough. The current study addressed the following questions: (1) Does increasing the mass of the femoral stem improve the taper connection strength intraoperatively? (2) Does increasing the mass of the femoral stem reduce the risk of periprosthetic tissue damage intraoperatively?

Hypothesis

Increasing the mass of the femoral stem improve the taper connection strength intraoperatively.

Materials and methods

During the experiment, femoral heads were impacted onto the stem tapers with and without an additional weight attached to the stem. The femoral heads were then pulled off to investigate the strength of the taper connection. The stem displacement and acceleration at impaction were also measured to evaluate the risk of periprosthetic tissue damage.

Results

The results showed that the pull-off force was increased by 24% (p=0.011, n=6) when an additional weight was attached to the stem. The additional weight also reduced the maximum stem acceleration and maximum stem displacement by 37% (p<0.001, n=6) and 14% (p=0.094, n=6), respectively.

Discussion

These findings suggest that the femoral head and stem taper connection strength can be significantly improved and the risk of periprosthetic tissue damage significantly reduced intraoperatively by attaching an additional weight to the stem to increase its mass.

Level of evidence

III, comparative in vitro mechanical investigation.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Total hip arthroplasty (THA), Impaction, Femoral head, Stem taper, Corrosion


Mappa


© 2022  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 108 - N° 8

Articolo 103284- dicembre 2022 Ritorno al numero
Articolo precedente Articolo precedente
  • Static and dynamic stabilometric force platform study of spinopelvic anterior sagittal imbalance
  • Mourad Ould-Slimane, François Luc, Nathalie Chastan, Franck Dujardin, Paul Michelin, Marie-Pierre Tavolacci, André Gillibert, Rémi Gauthé
| Articolo seguente Articolo seguente
  • Effect of screw tunnels on proximal femur strength after screw removal: A finite element analysis
  • Yu Zhang, An-An Li, Jia-Ming Liu, Wei-Lai Tong, Shi-Ning Xiao, Zhi-Li Liu

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2025 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.