Abstract

Background

Highly congruent total knee prostheses were introduced in the 1990s in the hope of decreasing polyethylene wear, thereby minimising loosening and particle-induced peri-prosthetic osteolysis. Despite promising long-term outcomes, substantial rates of aseptic loosening were reported with conventional gamma-irradiated polyethylene inserts, suggesting that highly reticulated polyethylene should be used instead. We assessed medium-term outcomes of the Wallaby I® total knee prosthesis with a deep-dished tibial insert made of conventional gamma-irradiated polyethylene.

Hypothesis

We hypothesised that the deep-dished Wallaby I® prosthesis was associated with similar or lower rates of aseptic loosening and peri-prosthetic osteolysis compared to posterior-stabilised prostheses.

Materials and methods

At our institution, 121 consecutive patients underwent total knee arthroplasty (TKA) with a deep-dished cemented prosthesis (Wallaby I®, Sulzer/Centerpulse, Zürich, Switzerland) between 2001 and 2005. Among them, 89 had complete follow-up data over a 4-year period and a mean follow-up of 96months. We retrospectively analysed the clinical and radiographic IKS scores in these 89 patients.

Results

Osteolysis with aseptic loosening required revision TKA of 10 knees after a mean follow-up of 81months. Mean 9-year prosthesis survival was 88±17%. Four inserts exhibited evidence of delamination. A fracture of the postero-medial aspect of the tibial baseplate beneath a zone of insert wear was found in 1 knee and gross mobility of the insert on the baseplate in 6 knees. The other 79 patients had good clinical and radiographic outcomes with a mean range of active knee flexion of 108±15°.

Discussion

The medium-term outcomes in our study were inferior to those reported with posterior-stabilised tibial components. Sporadic variations in polyethylene quality may explain the cases of osteolysis (shelf oxidation). In addition, the increased shear stresses related to the deep-dish design may increase backside wear, thereby compromising insert fixation to the baseplate. We believe the Wallaby I® prosthesis should no longer be used, and we recommend computed tomography follow-up of patients harbouring this prosthesis.

Level of evidence

Level IV (retrospective study).

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Deep-dished total knee arthroplasty, Particle-induced osteolysis, Loosening, Revision total knee arthroplasty


Mappa


© 2014  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 100 - N° 4

P. 403-408 - giugno 2014 Ritorno al numero
Articolo precedente Articolo precedente
  • Influence of the optical system and anatomic points on computer-assisted total knee arthroplasty
  • B. Schlatterer, J.-M. Linares, P. Chabrand, J.-M. Sprauel, J.-N. Argenson
| Articolo seguente Articolo seguente
  • Functional interest of an articulating spacer in two-stage infected total knee arthroplasty revision
  • T. Classen, M. von Knoch, M. Wernsmann, S. Landgraeber, F. Löer, M. Jäger

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2025 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.