Purpose of the study

Mechanical failure in total hip arthroplasty is usually related to aseptic loosening, itself related to wear particles as seen with polyethylene bearings. Alumina has been proposed to avoid wear problems. In vitro and mid-term clinical studies have shown tribological advantages but early acetabular fixation remains an issue. Since alumina-on-alumina bearings are currently used with new fixation techniques, updated evaluations of older series are informative regarding the long-term tolerance of alumina in vivo.

Material and methods

We investigated 104 consecutive alumina-on-alumina cemented total hip arthroàplasties (Cervaer-Osteal, Roissy, France) implanted 20 years ago in 81 patients (from 1979 to 1983) A 32-mm alumina femoral head was used. The alumina acetabular socket and the titanium femoral stem were cemented. The Postel-Merle-d'Aubigné score was used for clinical evaluation. Radiological wear and development of osteolysis or loosening were noted for establishing actuarial curves. When accessible, histological samples from revision procedures were analyzed.

Results

Six infected cases were not taken into account for data analysis. The average follow-up was 11 years, reaching 18 years in 38 cases. Twenty-five hips were revised for changing 23 acetabular sockets, 12 femoral heads, and one femoral stem. We noted one femoral head fracture, 24 definite acetabular loosenings, 12 probably acetabular loosenings, and 3 definite femoral loosenings. Radiological acetabular osteolysis was present in 4 cases, always limited to the DeLee zone 1, and associated with loosening. Radiological wear was below eye detection. Periprosthetic tissue showed non-specific histological reactions to cement particles. Survival rate at 20 years was 62.8% (49.3-76.2%) in terms of revision and 57.1% (42.571.7%) and 95.2% (89.9-100%) in terms of definite acetabular and femoral loosening.

Discussion

Besides the high rate of cemented fixation failure of the socket, loosened and non-loosened cases showed an excellent long-term tolerance of the alumina-on-alumina bearing with normal wear and osteolysis. This may also have protected the femoral component from complications. This study conrirms in vivo the long-term tribological benefit of the alumina-on-alumina bearing in total hip arthroplasty and suggests that acetabular fixation should be improved to achieve full benefit.

Keywords: Total hip arthroplasty , alumina , aseptic loosening , actuarial survival curve , long-term follow-up


Plan



© 2004 Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 90 - N° 8

P. 741-748 - décembre 2004 Retour au numéro

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.