S'abonner

Durom hip resurfacing system: Retrospective study of 644 cases with an average follow-up of 34 months - 04/04/13

Doi : 10.1016/j.otsr.2012.10.018 
S. Leclercq a, M. Lavigne b, J. Girard c, , P. Chiron d, P.A. Vendittoli b
a St Martin Hospital Center, 18, rue Rocquemonts, 14000 Caen, France 
b Maisonneuve-Rosemont Hospital, 5415, L’Assomption Boulevard, H1T 2M4, Montreal, Quebec, Canada 
c Lille Regional University Hospital Center, Lille University Sports Medicine Unit, Orthopaedics C Department, 2, avenue Oscar-Lambret, 59037 Lille cedex, France 
d University Hospital of Rangueil, 1, avenue J.-Poulhès, TSA 50034, 31059 Toulouse cedex 9, France 

Auteur correspondant. Tel.: +33 3 20 44 68 28; fax: +33 3 20 44 66 07.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Sous presse. Épreuves corrigées par l'auteur. Disponible en ligne depuis le Thursday 04 April 2013
Cet article a été publié dans un numéro de la revue, cliquez ici pour y accéder

Summary

Introduction

The performance of second-generation metal-on-metal bearings has led to the reintroduction of hip resurfacing. The goal of this multicentre study was to evaluate the short-term radiological and clinical outcomes with the Durom hip resurfacing system.

Hypothesis

The Durom hip-resurfacing system will have similar results to other hip resurfacing systems and traditional hip arthroplasty implants.

Materials and methods

In the four participating centers, 580 patients (406 men, 174 women) and 644 hips were included. The average patient age was 48 years (range 16–77). A posterolateral surgical approach was used in 357 cases; a Hardinge-type approach was used in 182 cases and a Rottinger-type approach in 105.

Results

After an average follow-up of 34 months, 31 hips (4.8%) had been revised. The reasons for revision were the following: 10 (1.6%) neck fracture (seven with Rottinger operative approach, one with Hardinge approach and one with posterolateral approach); 12 (1.9%) femoral loosening (four with lateral approach and eight with posterolateral approach); four (0.6%) acetabular cup migration; three (0.5%) unexplained pain; one (0.2%) adverse reaction to metal debris; one (0.2%) infection. Four hips (0.6%) dislocated but without recurrence – all were operated using the Hardinge approach. The 613 hips that were not operated on again had satisfactory clinical results; the Merle d’Aubigné score was 17.2 (range 12–18) and the WOMAC score was 91 (range 20–100). The five-year survival rate was 91% (95% CI: 87–94%). Based on radiographs, the average cup inclination was 44.4° (range 30 to 70°). The femoral offset was reduced by an average of 2.4mm (−31 to 23mm) and the leg length had increased by an average of 0.8mm (−15 to 19mm) relative to the other side, which was prosthesis-free. None of the non-revised implants showed any signs of loosening.

Discussion

This multicentre evaluation revealed that the Durom revision rate was slightly higher than the rate with other hip resurfacing systems and traditional total hip arthroplasty. Although the Durom system displayed excellent tribological performance, the differences relative to other implants may be attributed to the challenges associated with impaction, related to the geometry and design of the cup, and to precarious primary fixation. The choice of surgical exposure and implantation technique was an important factor in the survival of the implant.

Level of proof

Level IV – Retrospective study.

Le texte complet de cet article est disponible en PDF.

Keywords : Hip, Resurfacing, Revision, Biomechanical reconstruction, Arthroplasty


Plan


© 2013  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.