L’implant XEN® pour la prise en charge du glaucome chronique à angle ouvert : une revue de la littérature - 13/04/19

XEN® Gel Stent for management of chronic open angle glaucoma: A review of the literature (French translation of the article)

Doi : 10.1016/j.jfo.2018.10.005 
J. Buffault a, C. Baudouin a, b, c, A. Labbé a, b, c,
a Service d’ophtalmologie, centre hospitalier national d’ophtalmologie des Quinze-Vingts, 28, rue de Charenton, 75012 Paris, France 
b Service d’ophtalmologie, université de Versailles Saint-Quentin-en-Yvelines, hôpital Ambroise-Paré, AP–HP, 9, avenue Charles-De-Gaulle, 92100 Boulogne Billancourt, France 
c Inserm, U968, université UPMC Paris VI, UMR S 968, institut de la vision , CNRS, UMR 7210, 75012 Paris, France 

Auteur correspondant. Centre hospitalier national d’ophtalmologie des Quinze-Vingts - Service 3 , 28, rue de Charenton, 75012 Paris, France.France

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Résumé

Objectif

L’objectif de cette revue de la littérature était d’analyser l’effet de l’implant XEN® en procédure simple ou combinée à la phacoémulsification, sur la pression intraoculaire (PIO) et le nombre de traitements hypotonisants chez des sujets atteints de glaucome chronique à angle ouvert.

Méthode

Nous avons inclus les études de cohortes interventionnelles présentant au minimum un an de suivi postopératoire chez des patients atteints de glaucome primitif à angle ouvert (GPAO) ou secondaire à une pseudo-exfoliation capsulaire (GPEX) ou à une dispersion pigmentaire (GPIG) qui ont bénéficié de la pose d’un implant XEN®. Le critère de jugement principal était la réduction de la PIO à 12 mois postopératoires. Les critères de jugements secondaires étaient la baisse du nombre de traitements hypotonisants à 12 mois, la fréquence et le type de manœuvres postopératoires nécessaires et le taux de complications.

Résultats

Huit articles publiés entre 2016 et 2018 ont été inclus, six concernaient des études prospectives et deux, des études rétrospectives. Il n’y avait aucun essai contrôlé randomisé. Les données concernaient 958 yeux de 777 patients. Les différentes études retrouvaient une PIO à douze mois comprise entre 13 et 16mmHg en moyenne avec une réduction de PIO de 25 % à 56 % (moyenne : 42 %). Cette diminution était accompagnée d’une baisse du nombre de traitements hypotonisants dans toutes les études. La baisse de PIO à un an était significativement plus importante en procédure solo (44 %) qu’en procédure combinée (32 %) (p<0,05). Les complications les plus fréquentes étaient l’hypotonie transitoire (<1 mois) (3 %), l’apparition d’un décollement choroïdien ou de plis choroïdiens transitoires (1,5 %), un hyphéma (1,9 %), une fuite de la bulle de filtration (1,1 %) et une chambre antérieure étroite (1,1 %). Concernant les complications graves, quatre cas de glaucome malin (0,4 %) et un cas de décollement de rétine (0,1 %) ont été rapportés. En postopératoire, un needling a dû être réalisé dans 32 % des cas. Au total, 55 yeux (5,7 %) ont nécessité une nouvelle chirurgie filtrante ou la réalisation d’un cyclo-affaiblissement.

Conclusion

L’implant XEN® semble être efficace pour réduire la PIO et diminuer le nombre de traitements hypotonisants dans l’année suivant la chirurgie. Il présentait un profil de sécurité acceptable. Il nécessite cependant un suivi postopératoire rapproché et des manœuvres postopératoires fréquentes.

Le texte complet de cet article est disponible en PDF.

Summary

Purpose

The purpose of this study was to analyze the change in intraocular pressure (IOP) and glaucoma medications using the XEN® Gel Stent as a solo procedure or in association with phacoemulsification in patients with chronic open angle glaucoma (OAG).

Methods

We included cohort studies with at least one year of follow-up in patients with primary open angle glaucoma (POAG), pseudo-exfoliative glaucoma (PXG) or pigmentary glaucoma (PG) who received a XEN® gel stent. The main outcome measure was IOP reduction at 12 months follow-up. Secondary outcomes were the decrease in glaucoma medications 12 months after surgery, frequency and type of postoperative interventions and complication rate.

Results

A total of 8 case series published between 2016 and 2018 were included; six were prospective studies, and two were retrospective. There was no randomized controlled trial. The data concerned a total of 958 eyes of 777 patients. The various studies showed a mean IOP at 12 months between 13 and 16mmHg, which represented an IOP reduction between 25 and 56% (mean: 42%). This decrease was associated with a reduction in glaucoma medications in all studies. The decrease in IOP was significantly greater in XEN® implantation as a stand-alone procedure (44%) than in combined surgery (32%) (P<0.05). Transient hypotony (<1 month) (3%), choroidal detachment or choroidal folds (1.5%), hyphema (1.9%), bleb leak (1.1%) and shallow anterior chamber (1.1%) were the most frequent complications. As for severe complications, four cases of malignant glaucoma (0.4%) and one case of retinal detachment have been reported. In the follow-up period, needling has been required in 32% of cases, and a total of 55 eyes (5.7%) required repeat filtering surgery or cyclodestructive procedure.

Conclusion

XEN® Gel Stent appears effective for reducing IOP and the number of medications in OAG patients within 1 year postoperatively with an acceptable safety profile. However, vigilant postoperative follow-up and frequent postoperative maneuvers are required.

Le texte complet de cet article est disponible en PDF.

Mots clés : Glaucome, Glaucome chronique à angle ouvert, MIGS, Chirurgie du glaucome mini-invasive, Xen, Gel stent, Chirurgie du glaucome, Chirurgie combinée, Procédure solo, Bulle de filtration, Drainage sous-conjonctival, Pression intraoculaire

Keywords : Glaucoma, Chronic Open-Angle Glaucoma, MIGS, Minimally Invasive Glaucoma Surgery ;Xen, Gel Stent, Glaucoma Surgery, Combined Surgery, Solo procedure, Filtering Bleb, Subconjunctival drainage, Intraocular pressure


Plan


 Version française de l’article qui a été précédemment publié en anglais dans ce journal sous la référence: Buffault J, Baudouin C, Labbé A. XEN® Gel Stent for management of chronic open angle glaucoma: A review of the literature. J Fr Ophtalmol. 2019: doi: 10.1016/j.jfo.2018.12.002.
 French version of the article that has been previously published in English language in this journal under the reference: Buffault J, Baudouin C, Labbé A. XEN® Gel Stent for management of chronic open angle glaucoma: A review of the literature. J Fr Ophtalmol. 2019: doi: 10.1016/j.jfo.2018.12.002.


© 2019  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 42 - N° 4

P. 391-403 - avril 2019 Retour au numéro
Article précédent Article précédent
  • Prise en charge d’un carcinome épidermoïde de la conjonctive chez un sujet jeune : à propos d’un cas
  • S. Dib, R. Benhadouch, T. Abdou
| Article suivant Article suivant
  • Les bouchons lacrymaux : indications, efficacité et tolérance
  • A.-L. Best, M. Labetoulle, M. Legrand, M. M’garrech, E. Barreau, A. Rousseau

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.