Dual mobility hip arthroplasty wear measurement: Experimental accuracy assessment using radiostereometric analysis (RSA) - 04/10/10

Doi : 10.1016/j.otsr.2010.04.007 
V. Pineau a, , b , B. Lebel a, b, S. Gouzy a, b, J.-J. Dutheil b, c, C. Vielpeau a, b
a Department of Orthopaedic and Trauma Surgery, Caen “Côte de Nacre” Teaching Hospital, Center, avenue de la Côte-de-Nacre, 14033 Caen cedex France 
b Clinical Research in Orthopaedics Center (CROC), 14033 Caen, France 
c Clinical and Biostatistical Research Unit, Caen University Hospital Center, Côte de Nacre, 14033 Caen, France 

Corresponding author. Tel.: +33 2 31 06 51 26.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Summary

Introduction

The use of dual mobility cups is an effective method to prevent dislocations. However, the specific design of these implants can raise the suspicion of increased wear and subsequent periprosthetic osteolysis.

Hypothesis

Using radiostereometric analysis (RSA), migration of the femoral head inside the cup of a dual mobility implant can be defined to apprehend polyethylene wear rate.

Study objectives

The study aimed to establish the precision of RSA measurement of femoral head migration in the cup of a dual mobility implant, and its intra- and interobserver variability.

Material and methods

A total hip prosthesis phantom was implanted and placed under weight loading conditions in a simulator. Model-based RSA measurement of implant penetration involved specially machined polyethylene liners with increasing concentric wear (no wear, then 0.25, 0.5 and 0.75mm). Three examiners, blinded to the level of wear, analyzed (10 times) the radiostereometric films of the four liners. There was one experienced, one trained, and one inexperienced examiner. Statistical analysis measured the accuracy, precision, and intra- and interobserver variability by calculating Root Mean Square Error (RMSE), Concordance Correlation Coefficient (CCC), Intra Class correlation Coefficient (ICC), and Bland-Altman plots.

Results

Our protocol, that used a simple geometric model rather than the manufacturer’s CAD files, showed precision of 0.072mm and accuracy of 0.034mm, comparable with machining tolerances with low variability. Correlation between wear measurement and true value was excellent with a CCC of 0.9772. Intraobserver reproducibility was very good with an ICC of 0.9856, 0.9883 and 0.9842, respectively for examiners 1, 2 and 3. Interobserver reproducibility was excellent with a CCC of 0.9818 between examiners 2 and 1, and 0.9713 between examiners 3 and 1.

Discussion

Quantification of wear is indispensable for the surveillance of dual mobility implants. This in vitro study validates our measurement method. Our results, and comparison with other studies using different measurement technologies (RSA, standard radiographs, Martell method) make model-based RSA the reference method for measuring the wear of total hip prostheses in vivo.

Level of evidence

Level 3. Prospective diagnostic study.

El texto completo de este artículo está disponible en PDF.

Keywords : Radiostereometric Analysis (RSA), Total Hip Arthroplasty (THA), Wear, Dual Mobility


Esquema


© 2010  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 96 - N° 6

P. 609-615 - octobre 2010 Regresar al número
Artículo precedente Artículo precedente
  • Contents
| Artículo siguiente Artículo siguiente
  • Is alumina-on-alumina ceramic bearings total hip replacement the right choice in patients younger than 50 years of age? A 7- to 15-year follow-up study
  • P. Boyer, D. Huten, P. Loriaut, V. Lestrat, C. Jeanrot, P. Massin

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.