Reproducibility of an optical measurement system for the clinical evaluation of active knee rotation in weight-bearing, healthy subjects - 27/03/12

Doi : 10.1016/j.otsr.2011.08.017 
R. Testa a, b, c, , J. Chouteau a, b, c, d, A. Viste a, b, c, d, L. Cheze a, b, c, M.-H. Fessy a, b, c, d, B. Moyen a, b, c, d
a Lyon University, 69000, Lyon, France 
b IFSTTAR, UMR T9406, LBMC, Bron, France 
c Lyon 1 University, Villeurbanne, France 
d Orthopaedic Surgery Department, South Lyon-Hospital Center, 69495 Pierre-Bénite cedex, France 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Summary

Introduction

A knee is typically evaluated passively by a clinician during an office visit, without using dedicated measurement tools. When the knee is evaluated with the patient standing and actively participating in the movement, the results will differ than when the knee is passively moved through its range-of-motion by the surgeon. If a precise measurement system was available, it could provide additional information to the clinician during this evaluation.

Hypothesis

The goal of this study was to verify the reproducibility of a fast, flexible optical measurement system to measure rotational knee laxity during weight-bearing.

Material and methods

Two passive reflective targets were placed on the legs of 11 subjects to monitor femur and tibia displacements in three dimensions. Subjects performed internal and external rotation movements with the knee extended or flexed 30°. During each movement, seven variables were measured: internal rotation, external rotation and overall laxity in extension and 30° flexion, along with neutral rotation value in 30° flexion. Measurement accuracy was also assessed and the right and left knees were compared. Reproducibility was assessed over two measurements sessions.

Results

The calculated intra-class correlation coefficient (ICC) for reproducibility was above 0.9 for five of the seven variables measured. The calculated ICC for the right/left comparison was above 0.75 for five of the seven variables measured.

Discussion

These results confirmed that the proposed system provides reproducible measurements. Our right/left comparison results were consistent with the published literature. This system is fast, reproducible and flexible, which makes it suitable for assessing various weight-bearing movements during clinical evaluations.

Level of evidence

Level III, experimental study.

El texto completo de este artículo está disponible en PDF.

Keywords : Knee, Rotation, Laxity, Measurement system, Optical, Kinematics, Gait analysis


Esquema


© 2012  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 98 - N° 2

P. 159-166 - avril 2012 Regresar al número
Artículo precedente Artículo precedente
  • Accuracy of the preoperative planning for cementless total hip arthroplasty. A randomised comparison between three-dimensional computerised planning and conventional templating
  • E. Sariali, R. Mauprivez, F. Khiami, H. Pascal-Mousselard, Y. Catonné
| Artículo siguiente Artículo siguiente
  • Medial facet patelloplasty in patellar instability associated with patellofemoral dysplasia: A report of 26 cases
  • D. Saragaglia, R. Mader, M. Blaysat, N. Mercier

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.