Ulnar facet locking screw locations of volar locking plates placed without flexor pollicis longus tendon contact: A cadaver study - 10/04/20

Doi : 10.1016/j.otsr.2019.12.002 
Yuki Kikuchi a, Kotaro Sato a, , Yoshikuni Mimata a, Kenya Murakami a, Gaku Takahashi b, Minoru Doita a
a Department of Orthopaedic Surgery, Iwate Medical University, 1-1-1, Idaidori, Yahaba-cho, Shiwa-gun, 028-3694 Iwate, Japan 
b Department of Critical Care Medicine, Iwate Medical University, 1-1-1, Idaidori, Yahaba-cho, Shiwa-gun, 028-3694 Iwate, Japan 

Corresponding author at: Department of Orthopaedic Surgery, Iwate Medical University, 1-1-1, Idaidori, Yahaba-cho, Shiwa-gun, 028-3694 Iwate, Japan.Department of Orthopaedic Surgery, Iwate Medical University1-1-1, Idaidori, Yahaba-cho, Shiwa-gunIwate028-3694Japan

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Background

The volar locking plate (VLP) system provides stable fixation and is widely used for distal radius fractures. Studies have shown that irritation of the implant prominence is a risk factor for flexor tendon rupture, especially of the flexor pollicis longus (FPL). Contact between VLPs and flexor tendons should be avoided. We aimed to investigate the ulnar facet locking screw locations of various VLPs placed without FPL tendon contact in cadaver wrists.

Hypothesis

We hypothesized that ulnar facet locking screws would be in the vicinity of the subchondral bone when the plate is placed in the most distal position without FPL tendon contact.

Materials and methods

The study assessed two variable-angles and four fixed locking plates. We placed each plate in six different cadavers, resulting in 36 different plate-cadaver combinations. Plates were placed in the most distal position without FPL tendon contact. We drilled the most ulnar hole (hole A) and the second ulnar hole (hole B) of the distal row. All drilling procedures were performed using a specific jig for each fixed locking plate. For variable-angle locking plates, we drilled with a fixed jig for each plate. We obtained lateral radiographs when the drill penetrated the dorsal cortex and measured the distance between the drill and the articular surface.

Results

With regard to hole A, the mean distances between the drill and the center of the articular surface were 2.6–5.2mm for the four fixed locking plates and 4.9–5.6mm for the two variable-angle locking plates. With regard to hole B, the mean distances between the drill and the center of the articular surface were 3.8–5.9mm for the four fixed locking plates and 5.5–5.9mm for the two variable-angle locking plates.

Discussion

When clinicians place a VLP without FPL tendon contact, the distance between the ulnar facet screws and the center of the articular surface is over 3mm in most cases. Surgeons should select variable-angle drilling for strong articular support when using variable-angle locking plates.

Level of evidence

III, diagnostic Level.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Distal radius fracture, Volar locking plate, Flexor pollicis longus, Tendon rupture, Anatomic study


Mappa


© 2019  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 106 - N° 2

P. 365-370 - Aprile 2020 Ritorno al numero
Articolo precedente Articolo precedente
  • Chronological changes in surgical outcomes after trapeziectomy with ligament reconstruction and tendon interposition arthroplasty for thumb carpometacarpal osteoarthritis
  • Shingo Komura, Akihiro Hirakawa, Tomihiro Masuda, Marie Nohara, Ayaka Kimura, Yasuharu Matsushita, Haruhiko Akiyama
| Articolo seguente Articolo seguente
  • Treatment of supracondylar humeral non-union by bone autograft and Hoffmann II external fixation
  • Pieter Antoine Pierreux, Fabian Moungondo, Frédéric Arthur Schuind

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2024 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.