Right Parasternal Electrode Configuration Converts a Failed Electrocardiographic Screening to a Pass for Subcutaneous Implantable Cardioverter-Defibrillator Implantation - 18/11/15

Doi : 10.1016/j.hlc.2015.07.015 
N.Y. Chan, MBBS FRCP FACC FHRS , H.C. Yuen, MBBS, N.S. Mok, MBBS FRCP
 Department of Medicine & Geriatrics, Princess Margaret Hospital, Hong Kong 

Corresponding author at: Room 223, Block J, Princess Margaret Hospital, 2-10 Princess Margaret Hospital Road, Lai Chi Kok, Kowloon, Hong Kong. Tel.: +(852) 29903725; fax: +(852) 29903329.

Resumen

Background

Pre-operative electrocardiographic (ECG) screening before subcutaneous implantable cardioverter-defibrillator (SICD) implantation is essential to prevent T-wave oversensing and inappropriate shocks. The failure rate of ECG screening was reported to be up to 8% when only two body positions were tested.

Method

Three subcutaneous ECG vectors represented by lead I, II and III were obtained in standing, supine, sitting and squatting positions. A patient qualified if the ECG in any same lead passed in all four positions. We report a 31-year-old man with idiopathic ventricular fibrillation who failed ECG screening for SICD implantation with the conventional left parasternal electrode (LPS) configuration in all three subcutaneous ECG vectors. Right parasternal electrode (RPS) configuration with left arm and right arm ECG electrodes positioned 1cm lateral to right sternal border was attempted for screening.

Result

The amplitude of the QRS complex was significantly larger in the RPS compared to the LPS configuration in lead III and the patient passed the ECG screening in four body positions. He underwent successful SICD implantation with RPS approach with appropriate sensing both during the procedure and exercise treadmill test four weeks later. Ventricular fibrillation was successfully converted with 65J standard polarity shock during the procedure and no ICD shock was experienced by the patient on six-month follow-up.

Conclusion

RPS configuration may be considered in patients who fail the ECG screening with the conventional LPS approach for SICD implantation.

El texto completo de este artículo está disponible en PDF.

Keywords : Subcutaneous implantable cardioverter-defibrillator, Screening, Sudden cardiac death, Right parasternal, Subcutaneous electrocardiography



© 2015  Australian and New Zealand Society of Cardiac and Thoracic Surgeons (ANZSCTS) and the Cardiac Society of Australia and New Zealand (CSANZ). Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 24 - N° 12

P. e203-e205 - décembre 2015 Regresar al número
Artículo precedente Artículo precedente
  • Response to ‘Ex-vivo Donor Heart Perfusion: Testing the Limits of Cardiac Resilience’
  • N.L. Stamp, R.I. Larbalestier
| Artículo siguiente Artículo siguiente
  • Retrograde Type A Aortic Dissection Treated with Continuous Perfusion “Branch-First” Aortic Arch Replacement Technique
  • Louise C. Robertson, Damien Holdaway, Cheng-Hon Yap

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

¿Ya suscrito a @@106933@@ revista ?

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.