Dose reduction with iterative reconstruction in multi-detector CT: What is the impact on deformation of circular structures in phantom study? - 10/02/16

Doi : 10.1016/j.diii.2015.06.019 
J. Greffier a, , F. Macri a, A. Larbi a, A. Fernandez a, F. Pereira a, C. Mekkaoui a, b, 1, J.-P. Beregi a, 1
a EA 2415, Department of Radiology, Nîmes University Hospital, boulevard Professeur-Robert-Debré, 30029 Nîmes cedex, France 
b Martinos Centre for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, Boston, United States of America 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Objectives

To evaluate the distortion of circular structures induced by the increased image noise related to dose reduction and to assess the effect of iterative reconstruction (IR).

Methods

MDCT acquisitions were performed with 120 kVp for 200/100/60/40/20mAs with 100%/50%/30%/20%/10% of dose. Raw data were reconstructed by filtered back projection (FBP) and with two IR strengths. Image quality indices referred to water and acrylic were measured on a quality image phantom. Areas, perimeters, circularity were measured on the circular inserts of 4.8, 7.9 and 11.1mm on a morphological phantom.

Results

Dose reduction resulted in increased image noise and in decreased signal to noise ratio and contrast to noise ratio. IR improved these indices for the same dose without affecting the signal (number CT) and spatial resolution (modulation transfer function). The values of area, perimeter and circularity were altered compared to the actual value and the inserts were visually deformed with the dose reduction. IR improved these three parameters. Image quality indices, areas, perimeters and circularity of inserts were similar between the acquisition at 100% of the dose in FBP, 50% in strength-3 and 30% in strength-5 with different curves of noise power spectrum.

Conclusion

IR associated to 70% of dose reduction modifies the images smooth (NPS) but maintains adequate image quality indices without causing distortions of circular structures.

El texto completo de este artículo está disponible en PDF.

Keywords : Computed tomography, Iterative reconstruction, Image quality, Deformation of circular structures, Noise power spectrum

Abbreviations : CNR, CTDI, FBP, FOV, IR, MDCT, MTF, NPS, PSF, ROI, SAFIRE, SNR, TTF


Esquema


© 2015  Éditions françaises de radiologie. Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 97 - N° 2

P. 187-196 - février 2016 Regresar al número
Artículo precedente Artículo precedente
  • Efficacy of diffusion-weighted magnetic resonance imaging in the diagnosis and staging of endometrial tumors
  • I.S. Kececi, M.S. Nural, K. Aslan, M. Danac?, M. Kefeli, M. Tosun
| Artículo siguiente Artículo siguiente
  • The short- and long-term effectiveness of transcatheter arterial embolization in patients with intractable hematuria
  • M. Korkmaz, B. ?anal, B. Aras, H. Bozkaya, C. Ç?nar, S. Güneyli, M. Gök, G. Adam, F. Düzgün, ?. Oran

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.