Suscribirse

Electroanatomic Mapping and Transoesophageal Echocardiography for near Zero Fluoroscopy during Complex Left Atrial Ablation - 28/05/16

Doi : 10.1016/j.hlc.2016.01.018 
Hariharan Raju, MBChB, MRCP, PhD , John Whitaker, BMBCh, MRCP, Carly Taylor, BSc, Matthew Wright, MBBS, MRCP, PhD, FHRS
 Cardiology Department, St Thomas’ Hospital, London, UK 

Corresponding author at: Cardiac Electrophysiology Fellow / Registrar, Cardiology, East Wing Level 6, St Thomas’ Hospital, London SE1 7EH, United Kingdom. Tel.: +44 (0)207 1887335; fax: +44 (0)2071881011

Resumen

Background

We evaluated Carto 3, transoesophageal echocardiography (TOE) and contact force (CF) sensing catheter in atrial fibrillation (AF) ablation.

Methods

Unselected consecutive ablations performed under general anaesthesia (GA) were studied with modified protocol (cases, n=11) and compared to retrospective consecutive controls (n=10). Patent foramen ovale (PFO) or single transseptal puncture enabled left atrial (LA) access; ablation strategy was stepwise approach. Modified protocol utilised right atrial (RA) electrograms, CF and TOE to localise SmartTouch and create RA and CS electroanatomic map (EAM) without fluoroscopy. Transseptal puncture was performed with fluoroscopy in absence of PFO. Fluoroless pulmonary vein and LA EAM was created using TOE to locate circular-mapping catheter.

Results

Mean age of cases was 57±11 years with 64% male compared with 60±11 (70% male) for controls. Patent foramen ovale was identified in four cases (36%) and two controls (20%). No early complications occurred. Shorter fluoroscopy time (median 36 vs 390seconds; p=0.038) and trend to lower radiation dose (median 17 vs 165 cGym2; p=0.053) was seen in cases, with no increase in total procedure time (p=0.438).

Conclusions

General anaesthesia, TOE and CF mapping catheters facilitate minimised fluoroscopy for catheter ablation of LA arrhythmias. Zero fluoroscopy is possible in a majority of cases with PFO.

El texto completo de este artículo está disponible en PDF.

Keywords : Fluoroscopy, Radiation dosage, Catheter ablation, Atrial fibrillation, Foramen ovale, Patent


Esquema


© 2016  Australian and New Zealand Society of Cardiac and Thoracic Surgeons (ANZSCTS) and the Cardiac Society of Australia and New Zealand (CSANZ). Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 25 - N° 7

P. 652-660 - juillet 2016 Regresar al número
Artículo precedente Artículo precedente
  • Persistent Atrial Fibrillation Ablation using the Tip-Versatile Ablation Catheter
  • Edward J. Davies, Ben Clayton, Ian Lines, Guy A. Haywood
| Artículo siguiente Artículo siguiente
  • Transcatheter versus Surgical Aortic Valve Replacement in High-risk Patients: A propensity-score matched analysis
  • India Zweng, William Y. Shi, Sonny Palmer, Andrew MacIsaac, Robert Whitbourn, Philip Davis, Andrew E. Newcomb

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

¿Ya suscrito a @@106933@@ revista ?

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.