Suscribirse

Total knee implant posterior stabilised by a third condyle: Design evolution and post-operative complications - 01/12/16

Doi : 10.1016/j.otsr.2016.08.015 
R. Gaillard, S. Lustig , A. Peltier, V. Villa, E. Servien, P. Neyret
 Service d’orthopédie, Albert-Trillat Center, hôpital de la Croix-Rousse, 103, grande rue de la Croix-Rousse, 69004 Lyon, France 

Corresponding author. Tel.: +06 22 01 61 25.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Background

Despite excellent long-term outcomes, posterior stabilisation by a third condyle continues to receive unwarranted criticism regarding patellar complications and instability.

Hypothesis

Complication rates with a tri-condylar posterior-stabilised implant are similar to those with other posterior-stabilised prostheses and have diminished over time due to improvements in prosthesis design.

Material and methods

Post-operative complications and revision rates were assessed retrospectively in a prospective cohort of 4189 consecutive patients who had primary total knee arthroplasty (TKA) using a tri-condylar posterior-stabilised implant (Wright-Tornier) and were then followed-up for at least 24 months. The analysis included 2844 knees. The prosthesis generations were HLS1®, n=20; HLS2®, n=220; HLS Evolution®, n=636; HLS Noetos®, n=1373; and HLS KneeTec®, n=595. Complications were compared across generations by applying Fisher's exact test, and survival was compared using the Kaplan-Meier method.

Results

At last follow-up, there had been 341 (12%) post-operative complications in 306 (10.8%) knees, including 168 (5.9%) related to the implant, 41 (1.4%) infections, and 132 (4.6%) secondary complications unrelated to the implant. Re-operation was required for 200 complications (7%), including 87 (3.1%) consisting in revision of the prosthesis. Implant-related complications were stiffness (n=67, 2.4%), patellar fracture (n=34, 1.2%), patellar clunk syndrome (n=25, 0.9%), patellar loosening (n=3, 0.1%), tibial/femoral loosening (n=15, 0.5%), polyethylene wear (n=3, 0.1%), and implant rupture (n=1, 0.04%). Significant differences across generations were found for stiffness (P<0.0001), patellar fracture (P=0.03), clunk syndrome (P=0.03), and polyethylene wear (P=0.004), whose frequencies declined from one generation to the next. Overall 10-year survival was 92% with no significant difference across generations (P=0.1).

Discussion

Outcomes of tri-condylar posterior-stabilised TKA are similar to those obtained using other posterior-stabilised implants. Neither patellar complications nor instability are more common, and improvements in implant design have contributed to correct early flaws.

Level of evidence

IV, historical cohort, retrospective assessment of prospectively collected data.

El texto completo de este artículo está disponible en PDF.

Keywords : Total knee arthroplasty, Complications, Posterior stabilisation, Third midline condyle, Implant design


Esquema


© 2016  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 102 - N° 8

P. 1061-1068 - décembre 2016 Regresar al número
Artículo precedente Artículo precedente
  • Validation of a French patient-reported outcome measure for patello-femoral disorders: The Lille Patello-Femoral Score
  • S. Putman, F. Rémy, G. Pasquier, F. Gougeon, H. Migaud, A. Duhamel
| Artículo siguiente Artículo siguiente
  • Tibiofemoral syndesmosis injury treated by temporary screw fixation and ligament repair
  • S. Steinmetz, B. Puliero, D. Brinkert, N. Meyer, P. Adam, F. Bonnomet, M. Ehlinger

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.