Does using a polyethylene RM press-fit cup modify the preparation of the acetabulum and acetabular offset in primary hip arthroplasty? - 18/08/17

Doi : 10.1016/j.otsr.2017.03.026 
R. Erivan a, , S. Aubret b, G. Villatte a, A. Mulliez c, S. Descamps a, S. Boisgard a
a CNRS, SIGMA Clermont, ICCF, université Clermont Auvergne, CHU de Clermont-Ferrand, 63000 Clermont-Ferrand, France 
b Université Clermont Auvergne, CHU de Clermont-Ferrand, 63000 Clermont-Ferrand, France 
c Délégation à la recherche clinique et aux innovations (DRCI), CHU de Clermont-Ferrand, 63000 Clermont-Ferrand, France 

Corresponding author. Orthopedic and trauma surgery department, hôpital Gabriel-Montpied, CHU de Clermont-Ferrand, BP 69, 63003 Clermont-Ferrand, France.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Introduction

When performing total hip arthroplasty (THA), it is important to maintain the femoral and acetabular offsets to ensure good joint stability and to restore the function of the hip abductor muscles. In our practice, we mainly use a lateralized stem and hollow out the acetabulum to the quadrilateral plate to accommodate a press-fit polyethylene cup. However, the repercussions of this preparation method, which is driven by the cup's design, are not known. We carried out a retrospective study to assess: (1) the changes in the femoral and acetabular offset; (2) the height of the center of rotation; and (3) the repercussions on wear.

Hypothesis

We hypothesized there would be no significant differences between the preoperative and postoperative femoral and acetabular offsets.

Patients and methods

We reviewed 88 primary THA cases performed with the RM Pressfit™ cup that had a minimum of 5 years’ follow-up. A lateralized self-locking Muller-type cemented femoral stem was used in 92.0% of cases and a standard stem in 8.0%. Measurements were done on plain radiographs with MHP™ and Mesurim Pro™ software. The average follow-up was 6.5 years (5–8).

Results

On average, the acetabular offset was reduced by 2.75mm±5.9 mm (range: –17.5 to +10.6 mm) (P<0.001) and the femoral offset was increased by 0.01mm±5.5 mm (range: –17.8 to +11.0 mm) (P=0.99). In terms of total offset, medialization of 2.74mm±7 mm (range: –17.7 to +18.2mm) was found (P=0.001). The acetabular center of rotation was on average 4.77mm±5.1 mm higher (P<0.001). The mean annual wear at the more recent follow-up (min.: 5 years) was 0.068mm (range: 0.01 to 0.25mm) per year. The wear was not impacted by having more than 5mm change in offset.

Discussion

Measurements of acetabular offset revealed statistically significant medialization due to the type of implant used and the surgical technique. The anatomical technique consists of positioning the cup in subchondral bone without contacting the quadrilateral plate. This preserves bone stock, which may be useful later on if the cup is revised, particularly in younger patients. Conversely, the femoral offset did not change significantly, despite the use of lateralized stems in 92.0% of cases. We measured an annual wear rate of 0.068mm per year, which is lower than in other published studies, possibly because our patient population was older.

Level of evidence

IV, retrospective study.

El texto completo de este artículo está disponible en PDF.

Keywords : Acetabular offset, Femoral offset, Wear, Total offset, Press-fit polyethylene, RM cup, Total hip arthroplasty


Esquema


© 2017  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 103 - N° 5

P. 669-674 - septembre 2017 Regresar al número
Artículo precedente Artículo precedente
  • What are the risk factors for dislocation in primary total hip arthroplasty? A multicenter case-control study of 128 unstable and 438 stable hips
  • M.H. Fessy, S. Putman, A. Viste, R. Isida, N. Ramdane, A. Ferreira, A. Leglise, B. Rubens-Duval, N. Bonin, F. Bonnomet, A. Combes, S. Boisgard, D. Mainard, S. Leclercq, H. Migaud, SFHG 1
| Artículo siguiente Artículo siguiente
  • High-impact sport after hip resurfacing: The Ironman triathlon
  • J. Girard, A. Lons, T. Pommepuy, R. Isida, K. Benad, S. Putman

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.