Détermination expérimentale des facteurs de correction de quatre détecteurs actifs utilisés pour la dosimétrie des minifaisceaux utilisés en radiothérapie - 28/02/18

Experimental determination of correction factors of four detectors used in small field radiotherapy

Doi : 10.1016/j.canrad.2017.08.110 
L. Ben Salem ⁎ , A. Essadok, I. Saidani, M. Mahdouani, M. Benna, K. Mahjoubi, M. Besbes, F. Benna
 Unité de radiophysique, service de radiothérapie, institut Salah-Azaiz de Tunis, boulevard 9-Avril, 1006 Tunis, Tunisie 

⁎Auteur correspondant.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

páginas 7
Iconografías 3
Vídeos 0
Otros 0

Résumé

Objectif de l’étude

L’objectif de ce travail était la détermination expérimentale des facteurs de correction   de quatre dosimètres actifs de commerce : deux chambres d’ionisation et deux diodes solides sur la base des facteurs d’ouverture du collimateur de radiothérapie mesurés avec un film radiochromique pour un accélérateur linéaire équipé de cônes circulaires.

Matériel et méthodes

Dans un premier temps, un protocole de mesure de la dose par film radiochromique Gafchromic™ EBT3 avec une précision de 2 % a été élaboré pour approcher le « facteur d’ouverture du collimateur de référence ». Par la suite, les facteurs correctifs de quatre détecteurs ont été déterminés pour deux chambres d’ionisation (PinPoint PTW 31016 3D, Micropoint Extradin A16) et deux diodes solides (PTW T60017 Diode, PTW-60019 Micro-Diamond). Ces mesures ont été effectuées sous des collimateurs coniques de BrainLAB® définissant des faisceaux circulaires de diamètres à l’isocentre égaux à 7,5mm, 10mm, 12,5mm, 15mm, 17,5mm, 20mm, 25mm, 30mm, 35mm et 45mm d’un faisceau de photons X de 6MV généré par l’accélérateur linéaire ClinaciX (Varian®). Ces facteurs sont faiblement dépendants du type de l’accélérateur que ce soit le modèle ou le type collimation. Cela a permis leurs comparaisons avec ceux publiés pour le même type de détecteur et pour un accélérateur avec le même indice de qualité de faisceau.

Résultats

Les valeurs des facteurs correctifs, obtenues expérimentalement ont été comparables dans écart maximum de 1,9 % avec celles publiées des travaux utilisant le même type de détecteur (marque et modèle) et un accélérateur délivrant un faisceau de même indice de qualité pour la même taille de faisceau au point de mesure.

Conclusion

Le protocole de mesure par le film EBT3 utilisé comme dosimètre passif permettant de déterminer le « facteur d’ouverture du collimateur de référence » a été validé en comparant les données mesurées et publiées des facteurs de correction des détecteurs actifs.

El texto completo de este artículo está disponible en PDF.

Abstract

Purpose

The aim of this work is to determine experimentally the correction factors   for four active commercial dosimeters: two microchambers and two diode detectors based on the output factor measured with radiochromic film for a radiotherapy linear accelerator equipped with circular cones.

Materials and methods

Initially, a radiochromic film dosimetry measurement protocol with an accuracy of 2% was developed to approach the “reference output factor”. Afterwards, the corrective factors of four detectors were determined for two ionization chambers (PinPoint PTW 31016 3D, Micropoint Extradin A16) and two diodes (PTW T60017 Diode, PTW-60019 Micro-Diamond). These measurements were carried out under conical BrainLAB® collimators defining circular fields with diameters equal to 7.5mm, 10mm, 12.5mm, 15mm, 17.5mm, 20mm, 25mm, 30mm, 35mm and 45mm of a 6MV X-ray beam generated by the ClinaciX linear accelerator (Varian®). These factors are weakly dependent on the type of accelerator, whether the model and the collimation type. This allowed their comparisons with those published for the same type of detector and for an accelerator with the same index of beam quality.

Results

The correction factors obtained experimentally were comparable in maximum deviation of 1.9% with published ones of the works using the same type of detector (mark and model) and an accelerator delivering the same beam quality for the same field size at the measurement point.

Conclusion

The measurement protocol using the EBT3 film, which was used as a passive dosimeter to determine the “reference output factor”, was validated by comparing measured and published data of active detector correction factors.

El texto completo de este artículo está disponible en PDF.

Mots clés : Dosimétrie, Film, Gafchromic EBT3, Accélérateur linéaire

Keywords : Film, Dosimetry, Gafchromic EBT3, Linear accelerator


Esquema


© 2017  Société française de radiothérapie oncologique (SFRO). Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 22 - N° 1

P. 45-51 - février 2018 Regresar al número
Artículo precedente Artículo precedente
  • Outcome of postmastectomy radiotherapy after primary systemic treatment in patients with clinical T1-2N1 breast cancer
  • L. Cao, D. Ou, K.-W. Shen, G. Cai, R. Cai, F. Xu, S.-G. Zhao, C. Xu, N. Grellier Adedjouma, Y.M. Kirova, J.-Y. Chen
| Artículo siguiente Artículo siguiente
  • Investigation of electron boost radiotherapy in patients with breast cancer: Is a direct electron field optimal?
  • M. Aghili, M. Barzegartahamtan, A. Alikhassi, R. Mohammadpour

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
La compra de artículos no está disponible en este momento.

¿Ya suscrito a @@106933@@ revista ?

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.