Comparison of noise-magnitude and noise-texture across two generations of iterative reconstruction algorithms from three manufacturers - 28/06/19

Doi : 10.1016/j.diii.2019.04.006 
J. Greffier , A. Larbi, J. Frandon, G. Moliner, J.P. Beregi, F. Pereira
 Department of Radiology, CHU Nîmes, University Montpellier, Medical Imaging Group Nîmes, 30029 Nîmes, France 

Corresponding author at: CHU de Nîmes, Medical Imaging Group Nîmes, EA 2415, Bd Prof Robert Debré, 30029 Nîmes Cedex 9, France.CHU de Nîmes, Medical Imaging Group Nîmes, EA 2415Bd Prof Robert DebréNîmes Cedex 930029France

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Purpose

To compare the noise-magnitude and noise-texture across two generations of iterative reconstruction (IR) algorithms proposed by three manufacturers according to the dose level.

Materials and Methods

Five computed tomography (CT) systems equipped with two generations of IR algorithms (hybrid/statistical IR [H/SIR] or full/partial model-based IR [MBIR]) were compared. Acquisitions on Catphan 600 phantom were performed at 120kV and three dose levels (3-, 7- and 12-mGy). Raw data were reconstructed using standard “soft tissue” kernel for filtered back projection and one iterative level of two generations of IR algorithms. Contrast to-noise-ratio (CNR) was computed using three regions of interest: two of them placed in the low-density polyethylene (LDPE) and Teflon® inserts and another placed on the solid water. Noise power spectrum (NPS) was computed to assess the noise-magnitude (NPS peak) and noise-texture (NPS spatial frequency).

Results

CNR increased significantly in MBIR compared to H/SIR algorithms for General-Electric (GE) Healthcare (45%±12 [SD]) and Philips Healthcare systems (62%±11 [SD]) (P<0.001). Regarding Siemens Healthineers systems, CNR of MBIR was significantly lower than that of H/SIR (mean difference: −4%±5 [SD]) (P<0.001) for Teflon® insert but not for LDPE insert (mean difference: −4%±7 [SD]) (P=N.S.). NPS peaks were lower with MBIR than with H/SIR for GE Healthcare (-42%±8 [SD]) and Philips Healthcare (−75%±4 [SD]) systems, whereas it was greater with MBIR than with H/SIR for Siemens Healthineers (13%±11 [SD]) systems. NPS spatial frequencies were higher with MBIR than with H/SIR for Siemens (14%±10 [SD]) but lower for others (−17%±5 [SD] for GE Healthineers and −55%±3 [SD] for Philips Healthcare systems).

Conclusion

This study demonstrates that recent MBIR algorithms, by comparison with the preceding generation, differ according to the main manufacturers with respect to noise-magnitude and noise-texture.

El texto completo de este artículo está disponible en PDF.

Keywords : Multidetector computed tomography, Image quality enhancement, Iterative reconstruction, Optimization, Noise power spectrum

Abbreviations : CNR, CT, FBP, GE, H/SIR, IR, MBIR, NPS, ROI


Esquema


© 2019  Société française de radiologie. Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 100 - N° 7-8

P. 401-410 - juillet 2019 Regresar al número
Artículo precedente Artículo precedente
  • Interventional radiology techniques for pain reduction and mobility improvement in patients with knee osteoarthritis
  • D. Filippiadis, G. Charalampopoulos, A. Mazioti, E. Alexopoulou, T. Vrachliotis, E. Brountzos, N. Kelekis, A. Kelekis
| Artículo siguiente Artículo siguiente
  • Predicting TERT promoter mutation using MR images in patients with wild-type IDH1 glioblastoma
  • K. Yamashita, R. Hatae, A. Hiwatashi, O. Togao, K. Kikuchi, D. Momosaka, Y. Yamashita, D. Kuga, N. Hata, K. Yoshimoto, S.O. Suzuki, T. Iwaki, K. Iihara, H. Honda

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.