Suscribirse

Leadless Permanent Pacing: A Single Centre Australian Experience - 15/10/19

Doi : 10.1016/j.hlc.2018.09.014 
Russell A. Denman, MBBS a, ⁎ , Adam C. Lee, MBBS a, b, Caleb Mengel, MBBS a, b, Simon Townsend, BSc a, John Betts, BSc a, Nicole Bovey, BSc a, Daniel Wright, BSc a, Oscar Davison, BSc a, Haris M. Haqqani, MBBS, PhD a, b
a Department of Cardiology, The Prince Charles Hospital, Brisbane, Qld, Australia 
b Faculty of Medicine, University of Queensland, Brisbane, Qld, Australia 

⁎Corresponding author at: The Prince Charles Hospital, 627 Rode Rd, Chermside, QLD 4032.The Prince Charles Hospital627 Rode RdChermsideQLD4032

Resumen

Background

To describe the performance and clinical outcomes of consecutive patients having a leadless pacemaker (LP) implanted at a single institution.

Methods

Clinical data and device parameters were prospectively collected on all patients undergoing LP implantation from November 2015 to April 2018.

Results

A total of 79 patients (52 male), median age of 78 years, was included. Leadless pacemaker implantation was successful in 76 patients (96%). Implantation failed in two patients due to excessive venous tortuosity and due to inadequate sensing in another. Seventy-three (73) patients (96%) had chronic atrial fibrillation and all had a Class I or II indication for pacing. Procedure time was 29minutes (IQR 21–43) and fluoroscopy time was 8minutes (IQR 5–13). The median R wave at implant was 11.2mV (IQR 6.9–15.0). The median capture threshold at 0.24ms was 0.5V (IQR 0.4–0.9) and impedance was 754Ω (IQR 680–880). Intraprocedural acute dislodgement occurred in one patient following cutting of the tether but successful snaring and reimplantation was performed. During a median follow-up of 355days (range 9–905), overall electrical performance has been excellent. No patients have been readmitted for device revision or complications. Five (5) patients (7%) died during follow-up from unrelated causes.

Conclusions

Leadless pacemakers can be implanted safely and effectively in the majority of patients. Device electrical performance was excellent over a median follow-up of 12 months.

El texto completo de este artículo está disponible en PDF.

Keywords : Bradycardia, Heart block, Leadless pacemaker, Atrial fibrillation


Esquema


Crown Copyright © 2018  Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 28 - N° 11

P. 1677-1682 - novembre 2019 Regresar al número
Artículo precedente Artículo precedente
  • Assessment of the EuroSCORE II in a New Zealand Tertiary Centre
  • Navneet Singh, Damian Gimpel, Grant Parkinson, Paul Conaglen, Felicity Meikle, Zaw Lin, Nand Kejriwal, Nicholas Odom, David J. McCormack, Adam El-Gamel
| Artículo siguiente Artículo siguiente
  • Device-Related Thrombus After Left Atrial Appendage Occlusion With the Amulet Device
  • Michaël Peyrol, Jennifer Cautela, Erwan Salaun, Charlène Miola, Frederic Franceschi, Franck Thuny, Johan Pinto, Marc Laine, Alexandre Maximovitch, Franck Paganelli, Sébastien Armero, Laurent Bonello

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

¿Ya suscrito a @@106933@@ revista ?

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.