Comparative finite element analysis of three implants fixing stable and unstable subtrochanteric femoral fractures: Proximal Femoral Nail Antirotation (PFNA), Proximal Femoral Locking Plate (PFLP), and Reverse Less Invasive Stabilization System (LISS) - 30/01/20

Doi : 10.1016/j.otsr.2019.04.027 
Jie Wang a, b, Jian-Xiong Ma b, Bin Lu b, Hao-Hao Bai b, Ying Wang b, Xin-Long Ma a,
a Department of Orthopaedics, Tianjin Medical University General Hospital, No. 154, Anshan Street, Heping District, 300052, Tianjin China 
b Institute of Orthopaedics, Tianjin Hospital, No. 406, Jiefang Nan Street, Hexi District, 300211 Tianjin, China 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Background

The optimal type of fixation implant for managing subtrochanteric fractures (STFs) is debated, as uncertainty continues to surround the comparative biomechanical performance of the proximal femoral nail antirotation (PFNA), proximal femoral locking plate (PFLP), and reverse less invasive stabilisation system (LISS). No studies have used finite element analysis (FEA) to compare these three devices. The objective of this study was to use FEA to compare the PFNA, PFLP, and LISS used to treat STFs based on the following criteria: (1) stress distribution on the femur and implant, (2) peak stress and stress on the medial side of the femur near the fracture site, and (3) smallest axial displacement of the femoral head.

Hypothesis

Of the three implants, the PFNA has the best biomechanical performance when used for STF fixation.

Methods

FEA was used to assess synthetic bone responses to the three implants used to fix three STF types, namely, Seinsheimer I, III, and IV, characterised by increasing bone loss and/or comminution with subsequent instability. Loading was with 1400N axial compression force.

Results

The LISS and PFLP exhibited similar biomechanical properties in all three fracture types. However, with the Seinsheimer IV fracture, the triangular configuration of the PFLP resulted in stress concentration at the medial and lateral sides of the implant junction. With the Seinsheimer I and III fractures, the PFNA resulted in higher peak stress (183.85MPa and 364.58MPa, respectively) compared to the PFLP (102.90MPa and 177.52MPa) and LISS (116.55MPa and 227.97MPa). With the Seinsheimer IV fracture, peak stress was highest with LISS (2310.40MPa) and was higher with PFLP (2054.90MPa) than with PFNA (1313.30MPa). With the Seinsheimer I and III fractures, the axial femoral head displacement was greater with the PFNA (0.74mm and 1.13mm, respectively) than with the PFLP (0.48mm and 1.02mm) and LISS (0.52mm and 0.92mm). With the Seinsheimer IV fracture, in contrast, the PFNA produced less axial femoral head displacement (4.1mm) compared to the PFLP (12.03mm) and LISS (16.56mm).

Discussion

With unstable (Seinsheimer IV) STFs, fixation stability was better with the PFNA compared to the PFLP and LISS. In contrast, with stable STFs (Seinsheimer I and III), the PFLP and LISS offered greater stability, with similar biomechanical effects. However, with Seinsheimer III fractures, the stress on the implant-femur interface was greater with the LISS.

Level of evidence

IV, basic science study.

El texto completo de este artículo está disponible en PDF.

Keywords : Subtrochanteric fracture, Finite element analysis, Proximal femoral nail antirotation, Less invasive stabilisation system, Proximal femoral locking plate


Esquema


© 2019  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 106 - N° 1

P. 95-101 - février 2020 Regresar al número
Artículo precedente Artículo precedente
  • Application of the Radiographic Union Scale for Tibial fractures (RUST): Assessment of healing rate and time of tibial fractures managed with intramedullary nailing
  • Jun Min Leow, Nicholas D. Clement, A. Hamish W.R. Simpson
| Artículo siguiente Artículo siguiente
  • Annual case volume is a risk factor for 30-day unplanned readmission after open reduction and internal fixation of acetabular fractures
  • Robert Kent Merrill, Blake Ryan Turvey, Germanuel Lavar Landfair, Emmanuel Michael Illical

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.