Optimization of acquisition parameters for reduced-dose thoracic CT: A phantom study - 04/05/20

Doi : 10.1016/j.diii.2020.01.012 
K. Martini a, b, J.W. Moon a, M.P. Revel a, c, S. Dangeard a, C. Ruan d, G. Chassagnon a, c, e,
a Department of Radiology, Cochin Hospital, AP–HP Centre, 75014 Paris, France 
b Diagnostic and Interventional Radiology, University Hospital Zurich, 8008 Zurich, Switzerland 
c Université de Paris, Descartes-Paris 5, 75006 Paris, France 
d General Electric Healthcare, 78530 Buc, France 
e Center for Visual Computing, École Centrale Supelec, 91190 Gif-sur-Yvette, France 

Corresponding author at: Department of Radiology, Hôpital Cochin, 27, rue du Faubourg-Saint-Jacques, 75014 Paris, France.Department of Radiology, Hôpital Cochin27, rue du Faubourg-Saint-JacquesParis75014France

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Highlights

Favoring high tube voltage is the best option for reduced-dose CT protocols.
Lowest scores for subjective image quality are reached with low pitch factor.
Decrease of SNR does not necessarily go along with reduced subjective image quality.

El texto completo de este artículo está disponible en PDF.

Abstract

Purpose

The purpose of this study was to analyze the impact of different options for reduced-dose computed tomography (CT) on image noise and visibility of pulmonary structures in order to define the best choice of parameters when performing ultra-low dose acquisitions of the chest in clinical routine.

Materials and methods

Using an anthropomorphic chest phantom, CT images were acquired at four defined low dose levels (computed tomography dose index [CTDIvol]=0.15, 0.20, 0.30 and 0.40mGy), by changing tube voltage, pitch factor, or rotation time and adapting tube current to reach the predefined CTDIvol-values. Images were reconstructed using two different levels of iteration (adaptive statistical iterative reconstruction [ASIR®]-v70% and ASIR®-v100%). Signal-to-noise ratio (SNR) as well as contrast-to-noise ratio (CNR) was calculated. Visibility of pulmonary structures (bronchi/vessels) were assessed by two readers on a 5-point-Likert scale.

Results

Best visual image assessments and CNR/SNR were obtained with high tube voltage, while lowest scores were reached with lower pitch factor followed by high tube current. Protocols favoring lower pitch factor resulted in decreased visibility of bronchi/vessels, especially in the periphery. Decreasing radiation dose from 0.40 to 0.30mGy was not associated with a significant decrease in visual scores (P<0.05), however decreasing radiation dose from 0.30mGy to 0.15mGy was associated with a lower visibility of most of the evaluated structures (P<0.001). While image noise could be significantly reduced when ASIR®-v100% instead of ASIR®-v70% was used, the visibility-scores of pulmonary structures did not change significantly.

Conclusion

Favoring high tube voltage is the best option for reduced-dose protocols. A decrease of SNR and CNR does not necessarily go along with reduced visibility of pulmonary structures.

El texto completo de este artículo está disponible en PDF.

Keywords : Radiation dosage, Tomography, X-ray computed, Thorax, Phantoms


Esquema


© 2020  Publicado por Elsevier Masson SAS.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 101 - N° 5

P. 269-279 - mai 2020 Regresar al número
Artículo precedente Artículo precedente
  • COVID-19 pneumonia: A review of typical CT findings and differential diagnosis
  • C. Hani, N.H. Trieu, I. Saab, S. Dangeard, S. Bennani, G. Chassagnon, M.-P. Revel
| Artículo siguiente Artículo siguiente
  • Nodular reverse halo sign in active pulmonary tuberculosis: A rare CT feature?
  • K. Martini, A. Loubet, A. Bankier, S. Bouam, P. Morand, L. Cassagnes, M.-P. Revel, G. Chassagnon

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.