Optimization of image quality and accuracy of low iodine concentration quantification as function of dose level and reconstruction algorithm for abdominal imaging using dual-source CT: A phantom study - 04/01/22

Doi : 10.1016/j.diii.2021.08.004 
Djamel Dabli , Julien Frandon, Asmaa Belaouni, Philippe Akessoul, Takieddine Addala, Laure Berny, Jean-Paul Beregi, Joël Greffier
 Department of Medical Imaging, CHU Nîmes, Univ Montpellier, Medical Imaging Group Nimes, EA 2994, France 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Highlights

Dose reduction has an impact on image quality in virtual monochromatic images (VMIs) and iodine quantification using dual-energy CT.
Iterative reconstruction improves HU accuracy and image quality in VMIs and does not affect iodine accuracy.
Using ADMIRE for VMIs could help reduce radiation dose and/or the amount of injected iodine.

El texto completo de este artículo está disponible en PDF.

Abstract

Purpose

The purpose of this study was to assess the impact of advanced modeled iterative reconstruction (ADMIRE) algorithm and dose levels on the accuracy of Hounsfield unit (HU) measurement, image noise and contrast-to-noise ratio (CNR) in virtual monochromatic images (VMIs) with low iodine concentrations, and the accuracy of iodine quantification.

Materials and methods

A CT phantom was scanned with dual-source CT using abdomen-pelvis examination parameters at four dose levels: 5, 8, 11 and 20 mGy. Images were reconstructed using filtered-back projection (FBP) and ADMIRE levels 3 and 5 (A3-A5). HU accuracy was assessed calculating the root-mean-square deviation (RMSDHU). Image noise and CNR were computed on VMIs at 40/50/60/70 keV for 4 iodine inserts with 0.5, 1, 2 and 5 mg/mL concentrations. Accuracy of iodine quantification was assessed by the RMSDiodine and iodine bias (IB).

Results

The RMSDHU decreased significantly as the dose levels increased compared to 5 mGy, except for 8 mGy with A3 (P = 0.380) and with A5 level (P = 0.945). Noise increased by 63.0 ± 3.0 (standard deviation [SD])% from 20 mGy to 5 mGy. Noise decreased significantly by -53.8 ± 0.9 (SD) % with A5 compared to FBP. The CNR decreased by -43.1 ± 6.5 (SD)% from 20 mGy to 5 mGy. It increased using ADMIRE, and as the ADMIRE levels increased. The RMSDiodine and IB decreased as the dose level increased, and this was similar for all reconstruction types.

Conclusion

ADMIRE strongly improves image quality in VMIs and slightly improves HU accuracy but does not affect the accuracy of iodine quantification.

El texto completo de este artículo está disponible en PDF.

Keywords : Dual-energy CT, Iodine quantification, Monochromatic images, Iterative reconstruction

Abbreviations : ADMIRE, CNR, CT, CTDIvol, DECT, DLCT, DSCT, SECT, FBP, HU, IB, IR, MBIR, mAs, RMSD, SD, VMI


Esquema


© 2021  Société française de radiologie. Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 103 - N° 1

P. 31-40 - janvier 2022 Regresar al número
Artículo precedente Artículo precedente
  • Comparison of two deep learning image reconstruction algorithms in chest CT images: A task-based image quality assessment on phantom data
  • Joël Greffier, Julien Frandon, Salim Si-Mohamed, Djamel Dabli, Aymeric Hamard, Asmaa Belaouni, Philippe Akessoul, Francis Besse, Boris Guiu, Jean-Paul Beregi
| Artículo siguiente Artículo siguiente
  • Abdominal metastases from Merkel cell carcinoma: Prevalence and presentation on CT examination in 111 patients
  • Raphael Girard, Manel Djelouah, Maxime Barat, Paul Fornès, Sarah Guégan, Nicolas Dupin, Philippe Soyer, Christine Hoeffel

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.