Suscribirse

External validation of a commercially available deep learning algorithm for fracture detection in children - 24/02/22

Doi : 10.1016/j.diii.2021.10.007 
Michel Dupuis a, Léo Delbos b, Raphael Veil b, 1, Catherine Adamsbaum a, c, 1,
a AP-HP, Bicêtre Hospital, Pediatric Imaging Department, 94270 Le Kremlin Bicêtre, France 
b AP-HP, Bicêtre Hospital, Epidemiology and Public Health Department, 94270 Le Kremlin Bicêtre, France 
c Paris Saclay University, Faculty of Medicine, 94270 Le Kremlin Bicêtre, France 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Highlights

Deep learning algorithms lack real-world external validation prior to clinical use.
The tested deep learning algorithm shows strong diagnostic performance in children.
Sensitivity of the tested algorithm is lower in children under 4 years.

El texto completo de este artículo está disponible en PDF.

Abstract

Purpose

The purpose of this study was to conduct an external validation of a fracture assessment deep learning algorithm (Rayvolve®) using digital radiographs from a real-life cohort of children presenting routinely to the emergency room.

Materials and methods

This retrospective study was conducted on 2634 radiography sets (5865 images) from 2549 children (1459 boys, 1090 girls; mean age, 8.5 ± 4.5 [SD] years; age range: 0–17 years) referred by the pediatric emergency room for trauma. For each set was recorded whether one or more fractures were found, the number of fractures, and their location found by the senior radiologists and the algorithm. Using the senior radiologist diagnosis as the standard of reference, the diagnostic performance of deep learning algorithm (Rayvolve®) was calculated via three different approaches: a detection approach (presence/absence of a fracture as a binary variable), an enumeration approach (exact number of fractures detected) and a localization approach (focusing on whether the detected fractures were correctly localized). Subgroup analyses were performed according to the presence of a cast or not, age category (0–4 vs. 5–18 years) and anatomical region.

Results

Regarding detection approach, the deep learning algorithm yielded 95.7% sensitivity (95% CI: 94.0–96.9), 91.2% specificity (95% CI: 89.8–92.5) and 92.6% accuracy (95% CI: 91.5–93.6). Regarding enumeration and localization approaches, the deep learning algorithm yielded 94.1% sensitivity (95% CI: 92.1–95.6), 88.8% specificity (95% CI: 87.3–90.2) and 90.4% accuracy (95% CI: 89.2–91.5) for both approaches. Regarding age-related subgroup analyses, the deep learning algorithm yielded greater sensitivity and negative predictive value in the 5–18-years age group than in the 0–4-years age group for the detection approach (P < 0.001 and P = 0.002) and for the enumeration and localization approaches (P = 0.012 and P = 0.028). The high negative predictive value was robust, persisting in all of the subgroup analyses, except for patients with casts (P = 0.001 for the detection approach and P < 0.001 for the enumeration and localization approaches).

Conclusion

The Rayvolve® deep learning algorithm is very reliable for detecting fractures in children, especially in those older than 4 years and without cast.

El texto completo de este artículo está disponible en PDF.

Keywords : Radiographs, Deep learning algorithm, Artificial intelligence, Fractures, Pediatric

Abbreviations : CI, DICOM, FN, FP, LR-, LR+, NPV, PACS, PPV, ROI, TN, TP


Esquema


© 2021  Publicado por Elsevier Masson SAS.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 103 - N° 3

P. 151-159 - mars 2022 Regresar al número
Artículo precedente Artículo precedente
  • Local recurrence following complete radiologic response in patients treated with transarterial chemoembolization for hepatocellular carcinoma
  • Shamar Young, Tina Sanghvi, Sandeep Sharma, Cameron Richardson, Nathan Rubin, Masters Richards, Donna D'Souza, Siobhan Flanagan, Jafar Golzarian
| Artículo siguiente Artículo siguiente
  • Computed tomography angiography for quantification of cerebral vasospasm following aneurysmal subarachnoid hemorrhage
  • Mariam Soumah, Jonathan Brami, Davide Simonato, Benjamin Chousterman, Antoine Guillonnet, Anne-Laure Bernat, Emmanuel Houdart, Marc-Antoine Labeyrie

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.