Suscribirse

Évaluation de la performance clinique et analytique du système de mesure de glycémie Vital Check® MM-1200/bandelettes MS-2 dans la vraie vie - 11/10/23

Evaluation of the clinical and analytical performance in real life of the blood glucose monitoring system Vital Check® MM-1200/MS-2 strips

Doi : 10.1016/j.mmm.2023.04.002 
Amar Tebaibia 1, , Rachid Malek 2, Lakhdar Griene 3, Ali El Madhi Haddam 4, Nassim Nouri 5, Anwar Cherrak 6, Krim Belkacem 7, Samira Abrouk 8, Mourad Semrouni 9, Kamel Gasmi 10, Said Gari 10
1 Université Alger 1, établissement public hospitalier El Biar (Ex. Birtraria), service de médecine interne, El Biar, Alger, Algérie 
2 Université de Sétif, CHU de Sétif, service de médecine interne-diabétologie, Sétif, Algérie 
3 Université Alger 1, laboratoire d’hormonologie - centre Pierre et Marie Curie, Alger, Algérie 
4 Université Alger 1, CHU Bab El Oued, service de diabétologie-endocrinologie, Alger, Algérie 
5 Université de Constantine, CHU de Constantine, service d’endocrinologie-diabétologie, Constantine, Algérie 
6 Université d’Oran, CHU d’Oran, service de médecine interne, Oran, Algérie 
7 Maison du diabétique El Harrach, Alger, Algérie 
8 Méthodologie et Bio statistique consultante, Alger, Algérie 
9 Université Alger 1, CHU Beni-Messous, service de diabétologie-endocrinologie, Alger, Algérie 
10 Vital Care, production distribution et logistique, Alger, Algérie 

Amar Tebaibia, Établissement public hospitalier El Biar, service de médecine interne, 126 avenue Mustapha Ali Khodja, El Biar, Alger, Algérie.Établissement public hospitalier El Biar, service de médecine interne, 126 avenue Mustapha Ali KhodjaEl BiarAlgerAlgérie

Résumé

Introduction

La bonne gestion du diabète, les décisions thérapeutiques et la sécurité des patients dépendent de la performance analytique et clinique du système de mesure des glycémies capillaires. Celles-ci, effectuées par les patients, présentent des écarts par rapport aux glycémies veineuses mesurées au laboratoire.

Méthodes et objectifs

Étude nationale (Algérie), prospective, multicentrique, dans la vraie vie. L’objectif de l’étude était d’évaluer, dans la vraie vie, la concordance des résultats entre les glycémies capillaires mesurées par le système Vital Check® MM-1200 et les glycémies veineuses à jeun mesurées par un automate de laboratoire de biologie, chez des adultes atteints de diabète de type 1 (DT1) ou de diabète de type 2 (DT2). Les critères de l’exactitude analytique et de la performance clinique de la norme ISO 15197:2013 ont été évalués chez 462 patients DT1 ou DT2 issus de six centres.

Résultats

La concordance analytique globale était de 79,9 % [intervalle de confiance à 95 % : 79,9 ± 3,6] avec un coefficient de variation (CV) de 5,8 % et un biais moyen de -1,5 %. La performance clinique, en utilisant la grille d’erreur de Parkes, a objectivé que 99.2 % des valeurs étaient dans les zones A et B, pour lesquelles les différences entre glycémies capillaires et veineuses ne sont pas considérées comme déterminantes. Les facteurs limitant la performance de la mesure du glucose capillaire par le système Vital Check® MM-1200, sont liés aux patients, à l’environnement et à l’usage correct du système par les patients.

Conclusion

La concordance analytique du système de mesure de glycémie capillaire était sous-optimale. Elle était liée à des facteurs d’interférences. L’éducation thérapeutique du patient, pour la bonne pratique de l’autosurveillance glycémique (ASG), est essentielle pour limiter l’influence des facteurs exogènes sur les valeurs de glycémie capillaire mesurées qui pourraient avoir un impact négatif sur la décision clinique et thérapeutique.

El texto completo de este artículo está disponible en PDF.

Summary

Introduction

Successful diabetes management, treatment decisions, and patient safety depend on the analytical and clinical performance of capillary glucose monitoring systems. These, carried out by patients, show deviations from venous blood glucose levels measured in the laboratory.

Methods and aims

A national (Algerian), real-life, prospective, multicentric study. The objective of our study was to evaluate, in real life, the concordance between capillary glycemia measured by Vital Check® MM-1200 and fasting venous glycemia measured in the laboratory, in adults with type 1 or type 2 diabetes. The analytical accuracy criteria and the ISO 15197:2013 clinical performance were assessed in 462 adults with type 1 or type 2 diabetes, cared for in 6 centers.

Results

The overall analytical concordance was 79.9% [95% CI: 79.9 ± 3.6] with a coefficient of variation of 5.8% and a mean bias of -1.5%. The clinical performance, using the Parkes error grid, indicated that 99.2% of the values were in zones A and B, where differences between capillary and venous glycemia are not serious. Factors limiting the performance of the method of measuring capillary glucose were related to the patients, the environment and the proper use of this monitoring system.

Conclusion

Analytical concordance of this capillary blood glucose measurement system was suboptimal because of interfering factors. Therapeutic education, for the good practice of self-monitoring of blood glucose (SMBG), is essential to limit exogenous factors that influence the measured values of capillary glucose, that could have a negative impact on clinical and therapeutic decisions.

El texto completo de este artículo está disponible en PDF.

Mots clés : Autosurveillance glycémique, Vital Check® MM-1200, Concordance analytique, Performance clinique, Éducation thérapeutique, Algérie

Keywords : Self-monitoring of blood glucose, Vital Check® MM-1200, Analytical concordance, Clinical performance, Patient education, Algeria


Esquema


© 2023  Publicado por Elsevier Masson SAS.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 17 - N° 6

P. 507-516 - octobre 2023 Regresar al número
Artículo precedente Artículo precedente
  • Brèves
| Artículo siguiente Artículo siguiente
  • Impact de la mise en place d’une prise en charge psycho-nutritionnelle sur la qualité de vie des patients obèses
  • Elise Le Roy Feret, Estelle Caron, Jérémy Hermann, Anne-Laure Proust, Camille Gaulier, Erika Cornu, Anne-Cécile Paepegaey

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

¿Ya suscrito a @@106933@@ revista ?

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.