Validity and Reliability of 2 Cirrus Optical Coherence Tomography Procedures for Measuring Objective Ocular Torsion - 09/12/24

Doi : 10.1016/j.ajo.2024.08.004 
Elena Piedrahita-Alonso a, , Estela López-Redondo a, Alicia Valverde-Megías b, Gema Felipe-Márquez b, Beatriz Martín-García a, Rosario Gomez-de-Liano a, b
a From the Faculty of Optics and Optometry, Complutense University of Madrid, Madrid, Spain (E-P.A., E.L-R., B.M-G., R.G-L.) 
b Hospital Clínico San Carlos, Madrid, Spain (A.V-M., G.F-M., R.G-L.). 

Inquiries to Elena Piedrahita-Alonso, Complutense University of Madrid, Madrid, SpainComplutense University of MadridMadridSpain

Highlights

Two Cirrus optical coherence tomography (OCT) procedures measure ocular torsion with clinical advantages over fundus photography.
Objective ocular torsion by OCT does not require image export for further processing.
Ocular torsion assessment by the Cirrus OCT HD 1 line is fast, simple, and reliable.

El texto completo de este artículo está disponible en PDF.

Resumen

Purpose

Torsional eye position by the disc-fovea angle (DFA) is a relevant objective parameter in vertical strabismus. DFA measurement with optical coherence tomography (OCT) has proven to be a good alternative to the gold standard measurement in fundus photography. Our study aims to analyze the validity and reliability of 2 undescribed Cirrus high-definition OCT (HD-OCT; Carl Zeiss Meditec) methods that offer clinical advantages for measuring objective cycloposition in normal patients compared with Spectralis spectral-domain OCT (Heidelberg Engineering) and the reference method.

Design

Prospective validity and reliability analysis.

Methods

Objective cycloposition by means of the DFA was measured in the right eyes of 59 binocularly normal individuals attending the ophthalmology service of a medical institution. DFA was obtained by fundus photography (gold standard), FoDi software of Spectralis spectral-domain OCT, HD 1 line scan of Cirrus HD-OCT, and macular cube of Cirrus HD-OCT (fundus, FoDi, line, and cube methods, respectively). Measurements were performed 3 times for each method and the patient was repositioned and realigned between captures. Posterior manual quantification was made by 2 observers with external protractor software for the fundus and cube methods.

Results

The 3 OCT methods showed and excellent agreement with fundus photography (ICC 0.83-0.84) with no significant differences comparing mean values (P = .36 for fundus-FoDi, P = .09 for fundus-line, and P = .09 for fundus-cube). Absolute differences between methods were 1.5°. All methods showed excellent reliability (ICC 0.92 for FoDi, 0.91 for line, 0.92 for cube, and 0.91 for fundus). The minimal detectable change was lower than 3° and the absolute difference between repeated measurements was 1° for all methods. Interrater reliability was excellent for methods requiring manual quantification (ICC 0.98 for cube, ICC 0.94 for fundus).

Conclusions

Measurement of the DFA by Cirrus HD-OCT methods in normal patients was a valid and reliable alternative for the cycloposition assessment. Among the methods, the Cirrus OCT HD 1 line improved clinical performance due to the simplicity and speed of measurement, with no need to export the image for quantification.

El texto completo de este artículo está disponible en PDF.

Esquema


© 2024  The Author(s). Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 269

P. 273-281 - janvier 2025 Regresar al número
Artículo precedente Artículo precedente
  • Evaluation of a Novel Virtual Reality Simulated Alternate Cover Test to Assess Strabismus: A Prospective, Masked Study
  • DIEGO MARTINEZ MORI, ASHIKA KUCHHANGI, JESSICA TAME, KAREN COOPER, LEILA HAJKAZEMSHIRAZI, MAANASA INDARAM, JEREMY D. KEENAN, JULIUS T. OATTS
| Artículo siguiente Artículo siguiente
  • Performance Evaluation of a Simple Strategy to Optimize Formula Constants for Zero Mean or Minimal Standard Deviation or Root-Mean-Squared Prediction Error in Intraocular Lens Power Calculation
  • Achim Langenbucher, Nóra Szentmáry, Jascha Wendelstein, Alan Cayless, Peter Hoffmann, Damien Gatinel

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

¿Ya suscrito a @@106933@@ revista ?

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.