Improved myocardial scar visualization using free-breathing motion-corrected wideband black-blood late gadolinium enhancement imaging in patients with implantable cardiac devices - 06/05/25

Doi : 10.1016/j.diii.2024.12.001 
Pauline Gut a, b, c, Hubert Cochet d, Panagiotis Antiochos e, Guido Caluori a, Baptiste Durand a, Marion Constantin a, Konstantinos Vlachos a, Kalvin Narceau a, Ambra Masi e, Jürg Schwitter c, e, Frederic Sacher a, f, Pierre Jaïs a, f, Matthias Stuber a, b, g, Aurélien Bustin a, b, d,
a IHU LIRYC, Heart Rhythm Disease Institute, Université de Bordeaux, INSERM U1045, 33604, Pessac, France 
b Department of Diagnostic and Interventional Radiology, Lausanne University Hospital and University of Lausanne, 1011 Lausanne, Switzerland 
c Faculty of Biology and Medicine, University of Lausanne, 1011 Lausanne, Switzerland 
d Department of Cardiovascular Imaging, Hôpital Cardiologique du Haut-Lévêque, CHU de Bordeaux, 33604 Pessac, France 
e Cardiovascular Department, Division of Cardiology, University Hospital of Lausanne and University of Lausanne, 1011 Lausanne, Switzerland 
f Department of Cardiac Pacing and Electrophysiology, Hôpital Cardiologique du Haut-Lévêque, CHU de Bordeaux, 33604 Pessac, France 
g CIBM Center for Biomedical Imaging, 1011 Lausanne, Switzerland 

Corresponding author.

Highlights

Free-breathing wideband black-blood late gadolinium enhancement cardiac MR images are reconstructed in <2 s.
Non-rigid motion reconstruction enables motion-free black-blood late gadolinium enhancement cardiac MR imaging.
Free-breathing wideband black-blood late gadolinium enhancement MR imaging enables improved detection of myocardial scar and suppression of artifacts in the myocardium associated with implantable cardiac devices.

El texto completo de este artículo está disponible en PDF.

ABSTRACT

Purpose

The purpose of this study was to introduce and evaluate a novel 2D wideband black-blood (BB) LGE sequence, incorporating wideband inversion recovery, wideband T2 preparation, and non-rigid motion correction (MOCO) reconstruction, to improve myocardial scar detection and address artifacts associated with implantable cardioverter defibrillators (ICDs).

Materials and methods

The wideband MOCO free-breathing BB-LGE sequence was tested on a sheep with ischemic scar and in 22 patients with cardiac disease, including 15 with cardiac implants, at 1.5T. Wideband MOCO free-breathing BB-LGE sequence was compared with conventional and wideband breath-held PSIR-LGE and conventional and wideband breath-held BB-LGE techniques. Image sharpness, entropy, and scar-to-blood, scar-to-myocardium, and blood-to-myocardium contrast were analyzed and reconstruction times were measured. Two expert readers assessed the image quality, ICD artifact severity, and the diagnostic confidence with scar extent. Finally, for the animal study, a histology of the heart was performed to confirm the presence and localization of scar tissue.

Results

In the animal, wideband MOCO free-breathing BB-LGE were reconstructed in 0.6 s and demonstrated a 200 % improvement in scar-to-blood contrast compared to wideband breath-held PSIR-LGE, with significant improvement in image sharpness and reduction in entropy. It also effectively minimized ICD artifacts and accurately detected scars. In patients, wideband MOCO free-breathing BB-LGE were reconstructed in 1.5 ± 0.4 (standard deviation) s per slice. Seventeen patients (17/22; 77%) with myocardial scars were confidently diagnosed with wideband MOCO free-breathing BB-LGE, compared to 11 (11/22; 50 %) with wideband breath-held PSIR-LGE (P < 0.01).

Conclusion

Free-breathing wideband T2-prepared black-blood LGE imaging, combined with motion-corrected reconstruction, offers a promising diagnostic approach for the evaluation of myocardial lesions in patients with ICDs.

El texto completo de este artículo está disponible en PDF.

Keywords : Black blood imaging, Implantable cardioverter defibrillator, Cardiac magnetic resonance imaging, Motion artifacts, Myocardial infarction

Abbreviations : 2D, AHA, BB, BW, CI, CIED, CMR, CRT-D, GRAPPA, GRE, ICD, IR, LGE, LVEF, MOCO, MRI, N/A, NYHA, PSIR, Q1, Q3, RVEF, S-ICD, SA, SCA, SD, T2prep, TI, TV- ICD, VT


Esquema


© 2024  The Author(s). Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 106 - N° 5

P. 169-182 - mai 2025 Regresar al número
Artículo precedente Artículo precedente
  • Evaluation of navigation and robotic systems for percutaneous image-guided interventions: A novel metric for advanced imaging and artificial intelligence integration
  • Francois H. Cornelis, Dimitrios K Filippiadis, Philipp Wiggermann, Stephen B. Solomon, David C. Madoff, Laurent Milot, Sylvain Bodard
| Artículo siguiente Artículo siguiente
  • Real-time multislice MR-thermometry of the prostate: Assessment of feasibility, accuracy and sources of biases in patients
  • Clément Marcelin, Amandine Crombé, Eva Jambon, Grégoire Robert, Franck Bladou, Pierre Bour, Thibaut Faller, Valéry Ozenne, Nicolas Grenier, Bruno Quesson

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
El acceso al texto completo de este artículo requiere una suscripción.

¿Ya suscrito a @@106933@@ revista ?

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.