Intraoperative reference points on the proximal tibia in image-free robotic-assisted total knee arthroplasty should be determined by preoperative posterior tibial slope - 03/06/25

Doi : 10.1016/j.otsr.2025.104180 
Yutaka Nakamura, Hiroyasu Ogawa , Haruhiko Akiyama
 Department of Orthopaedic Surgery, Gifu University Graduate School of Medicine, Yanagido 1-1, Gifu, 501-1194, Japan 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Background

The purpose of this study was to investigate the accuracy of the tibial cut in sagittal plane and intraoperative optimal reference points on the proximal tibia for achieving the targeted posterior tibial slope (PTS) in image-free, robotic-assisted total knee arthroplasty (TKA).

Hypothesis

A mechanical tibial axis determined by intraoperative reference points would affect the measurement of the PTS and thereby postoperative PTS in image-free robotic assist TKA.

Patients and Methods

Fifty-eight patients (70 knees) who underwent primary image-free robotic-assisted TKA were included. Pre- and postoperative PTS were evaluated using whole-leg computed tomography images, which were analysed with three-dimensional planning software. Change in PTS (ΔPTS) was calculated by subtracting the target PTS from postoperative PTS. The proximal tibial axis ratio was defined as the proportion of distance between the anterior border of the anterior cruciate ligament (ACL) footprint and the tibial axis on the proximal joint surface to the anteroposterior width of the ACL footprint.

Results

The mean ΔPTS was −0.4 ± 2.0 °. Eight outliers (11.4%; |ΔPTS| >3°) were identified. The proximal tibial axis ratio was −13.2 ± 19.9% and showed a significant negative correlation with preoperative PTS and ΔPTS (r = −0.87 and −0.29, p < 0.001 and p = 0.01, respectively). The tibial axis passed through the anterior border of the ACL footprint when preoperative PTS was 9.6 °. These results indicated that a larger preoperative PTS was associated with a more anterior tibial axis on the proximal joint surface. Preoperative PTS significantly correlated with ΔPTS (r = 0.34 and p = 0.004).

Discussion

In image-free robotic-assisted TKA, when the preoperative PTS is >9.6 °, positioning the proximal tibial reference point anterior to the anterior border of the ACL footprint is recommended.

Level of evidence

III.1

El texto completo de este artículo está disponible en PDF.

Keywords : Posterior tibial slope, Reference point, Robotic-assisted total knee arthroplasty, Tibial axis


Esquema


© 2025  Publicado por Elsevier Masson SAS.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 111 - N° 4

Artículo 104180- juin 2025 Regresar al número
Artículo precedente Artículo precedente
  • Is pre-operative block-test with lidocaine injection efficient in predicting the functional result of revision total hip arthroplasty?
  • Maxime Wolf, Philippe-Alexandre Faure, Julien Dartus, Thomas Amouyel, Hélène Behal, Julien Girard, Henri Migaud, Pierre Martinot
| Artículo siguiente Artículo siguiente
  • Robot-assisted surgery and knee arthroplasty in genu valgum: impact of the surgical approach
  • Emeline Chapron, Rémy Coulomb, Aymeric Weiss, Philippe Marchand, Pascal Kouyoumdjian

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.