Total knee arthroplasty three-dimensional kinematic estimation prevision. From a two-dimensional fluoroscopy acquired dynamic model - 09/04/11

Doi : 10.1016/j.otsr.2011.01.003 
B.P. Lebel a, , V. Pineau a, S.L. Gouzy a, L. Geais b, J.-J.M. Parienti c, J.-J.P. Dutheil c, C.H. Vielpeau a
a Orthopedic Department, Caen University Hospital, avenue de la Côte-de-Nacre, 14033 Caen cedex, France 
b Reasearch and Development, Stryker SA, Europe 
c Biostatistic Department, Caen University Hospital, Caen, France 

Corresponding author. Tel.: +33 2 31 06 48 09; fax: +33 23 10 64 85 09.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Summary

Introduction

To determine six-degree of freedom of total knee arthroplasty kinematics (TKA), optimized matching algorithms for single fluoroscopic image system may be used. Theoretical accuracy of these systems was reported. Nevertheless, all reports were done under idealized laboratory experimental conditions. The aim of this study was to evaluate the “true” accuracy of a flat panel single plane video-fluoroscopy system based on computed-assisted design (CAD) model matching and compare it to TKA kinematics obtained from optoelectronic measurements as gold standard.

Hypothesis

The estimation of the error produced by 2D/3D fluoroscopic registration in daily practice is misjudged in most available laboratory reports.

Material and methods

The experimental set-up used a TKA implanted into femoral and tibial cadaver bones. Thirty flexions were simultaneously registered using single plane fluoroscopy and an active optical tracking system. Kinematics registered were compared using the root mean square error (RMS), the concordance correlation coefficient and Bland & Altman plot analysis.

Results

The mean range of motion of flexion during the experiment was 106°. The respective RMS for flexion, varus-valgus and internal-external rotation were 0.68, 0.67 and 1.02°. The respective RMS for antero-posterior, medio-lateral and proximo-distal displacement were 1.3, 2.4 and 1.06mm. Extreme values of the measured error concerning medio-lateral displacement were −5.4 and 22,1mm.

Discussions

Analysis found some outliners in all degree of freedom with a systematic error and larger standard deviation than already published data. One should make sure that during the experiment the motion of interest is in the in-plane direction. Moreover, this study brings out the true threshold detection of this type of analysis.

Level of evidence

Level 3. Prospective diagnostic study.

El texto completo de este artículo está disponible en PDF.

Keywords : Kinematics, Total knee arthroplasty, Fluoroscopy, Image processing, Optoelectronic, Accuracy


Esquema


© 2011  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 97 - N° 2

P. 111-120 - avril 2011 Regresar al número
Artículo precedente Artículo precedente
  • Contents
| Artículo siguiente Artículo siguiente
  • Stem subsidence after total hip revision: 183 cases at 5.9 years follow-up
  • J. Girard, O. Roche, G. Wavreille, F. Canovas, P. Le Béguec

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.