Dual mobility retentive acetabular liners and wear: Surface analysis of 40 retrieved polyethylene implants - 04/02/14

Doi : 10.1016/j.otsr.2013.12.011 
P. Adam , F. Farizon, M.-H. Fessy
 Centre d’orthopédie et traumatologie, hôpital Bellevue, CHU de Saint-Étienne, 42055 Saint-Étienne cedex 2, France 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Purpose of the study

Dual mobility systems with retentive acetabular liners have been used in clinical practice for total hip arthroplasty since 1976. The dual mobility system preserves a wide range of motion while providing greater stability. This study measured wear on the concave and convex surfaces of 40 retrieved polyethylene liners, to evaluate the advantages of this system in relation to wear.

Material and methods

Forty polyethylene inserts that had been removed due to infection or mechanical failure after a mean 8 years were analyzed. The mean age of patients at arthroplasty was 46 years old. Macroscopic analysis was followed by surface analysis with direct measurement of changes in the curvature radii. The internal concave surface was measured in three dimensions using a 4-mm stylus (BNH 706). External convexity was measured by lateral projection. The estimated error was ± 5 μm for both measurement methods. Manufacturers’ tolerance for these implants was approximately 50 μm. Linear wear and wear volume was determined by comparing the measured dimensions with the theoretical dimensions of new liners.

Results

Macroscopically, all of the pieces studied had lost the initial machined grooves on the convex surface; 40% of the pieces showed visible wear of the retentive collar. Mean annual convex surface wear was 9 μm (SD 9μm) and 73 μm (SD 69 μm) for the concave surface. Mean total wear, which was the sum of the wear on the convex and concave surfaces was 82 μm (SD 72 μm). Wear volume was 28.9 mm3/yr for the convex surface (SD 27.6) and 25.5 for the concave surface (SD 23.2) with a mean annual total wear volume of 54.3 mm3/yr (SD 39.6).

Discussion

Total wear in the 40 dual mobility liners that had functioned in vivo was similar to that reported in metal-polyethylene bearings with 22.2mm femoral heads. The results of wear in both the convex and concave surfaces show that wear with the dual mobility system was not increased compared to conventional metal-polyethylene bearings, while providing better retention and greater stability.

Conclusion

The use of dual mobility acetabular liners is an attractive solution when a metal-polyethylene bearing is needed. The increased joint stability is not associated with increased wear.

El texto completo de este artículo está disponible en PDF.

Keywords : Hip, Prosthesis, Dual Mobility, Bipomar prosthesis


Esquema


 Original article. For citation, use not the present reference but that of the original publication: Adam P, Farizon F, Fessy MH. Dual articulation retentive acetabular liners and wear: surface analysis of 40 retrieved polyethylene implants. Rev Chir Orthop Reparatrice Appar Mot 2005;91(7):627-36.


© 2013  Publicado por Elsevier Masson SAS.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 100 - N° 1

P. 85-91 - février 2014 Regresar al número
Artículo precedente Artículo precedente
  • Gamma nail in the treatment of closed trochanteric fractures. Results and indications of 121 cases
  • I. Kempf, A. Grosse, G. Taglang, E. Favreul
| Artículo siguiente Artículo siguiente
  • Grammont reversed prosthesis for acute complex fracture of the proximal humerus in an elderly population with 5 to 12 years follow-up
  • J.-F. Cazeneuve, D.-J. Cristofari

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.